II. Product description
2.1.3
Determination of the isoelectric point
Acid/ base titration
Apart from charge measurements, the Mütek™ PCD-06
lends itself to identifying the isoelectric point (IEP) of
dispersions.
The IEP is the pH-dependent point of zero charge of a
particle.
By adding a titrant (acid/ base) drop by drop, the
sample’s pH is shifted until the isoelectric point (strea-
ming potential = 0 mV) is reached.
Cationic samples are titrated with a base and anionic
samples with an acid up to the point of 0 mV.
Fig. 2-2
Schematic of isoelectric point
2.1.4
Titrations curves
Information
The Mütek™ PCD-06 Standard or Premium conti-
nuously generates a titration curve while performing a
sample titration.
Titration curves are basically difficult to interpret,
because the curve progression does not exclusively
reflect the sample material, but rather the adsorption
behaviour of the sample at the surface of the measuring
cell. This is why the curve progression should not be
taken as a basis for the discussion of results. Instead,
under certain conditions, the curves provide information
about the nature of the sample under test. The titration
curves shown in the following may roughly be associa-
ted with different stock systems:
2.1.4.1 Colloidally dissolved particles
A titration curve obtained for a colloidally dissolved
sample is shown in Fig. 2-3.
At the onset of titration, the streaming potential remains
essentially unchanged, since the non-adsorbed colloids
are neutralized and do not therefore contribute toward
the generation of a streaming potential.
Later on, the streaming potential undergoes a sharp
drop, which is indicative of the fact that adsorbed
charge carriers are neutralized at the cell wall and pis-
ton.
Fig. 2-3
Colloidally dissolved particles
IEP
Summary of Contents for BTG Mutek PCD-06
Page 1: ...Operation Manual M tekTM PCD 06 Light Standard Premium Particle Charge Detector...
Page 6: ...List of contents Chapter Paragraph Title Page 6 Operation Manual M tekTM PCD 06 BTG...
Page 20: ...20 Operation Manual M tekTM PCD 06 BTG 2021 II Product description...
Page 24: ...22 Operation Manual M tekTM PCD 06 BTG 2021 III System layout and functions...
Page 32: ...30 Operation Manual M tekTM PCD 06 BTG 2021 V Measuring 5 3 Overview Main Screens...
Page 56: ...54 Operation Manual M tekTM PCD 06 BTG 2021 VIII Shutting down the system storage and disposal...
Page 59: ......
Page 60: ...BTG Instruments AB P O Box 602 SE 661 29 S ffle Sweden Phone 46 533 426 00 www btg com...