INTRODUCTION

INTRODUCTION

CAUTION

* This device is intended for adult use in homes only. 
* The device is not suitable for use on neonatal patients, pregnant women, patients with 
implanted electronical devices, patients with pre-eclampsia, premature ventricular beats, 
atrial fibrillation, peripheral arterial disease and patients undergoing intravascular therapy or 
arterio-venous shunt or people who received a mastectomy. Please consult your doctor 
prior to using the unit if you suffer from an illnesses.
* The device is not suitable for measuring the blood pressure of children. Ask your doctor 
before using it on older children.
* The device is not intended for patient transport outside a healthcare facility.
* The device is not intended for public use.
* This device is intended for non-invasive measuring and monitoring of arterial blood 
pressure. It is not intended for use on extremities other than the arm or for functions other 
than obtaining a blood pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your 
blood pressure. Do not begin or end medical treatment without asking a physician for 
treatment advice.
* If you are taking medication, consult your physician to determine the most appropriate 
time to measure your blood pressure. Never change a prescribed medication without 
consulting your physician.
* Do not take any therapeutic measures on the basis of a self-measurement. Never alter the 
dose of a medicine prescribed by a doctor. Consult your doctor if you have any question 
about your blood pressure.
* When the device is used by patients who have common arrhythmias such as atrial or 
ventricular premature beats or atrial fibrillation, the best result may occur with deviation. 
Please consult your physician about the result.
* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously 
increase which can prevent blood flow and result in harmful injury to the PATIENT.
* When using this device, please pay attention to the following situation which may interrupt 
blood flow and influence blood circulation of the patient, thus cause harmful injury to the 
patient: connection tubing kinking too frequent and consecutive multiple measurements; the 
application of the cuff and its pressurization on any arm where intravascular access or 
therapy, or an arterio-venous (A-V) shunt, is present; inflating the cuff on the side of a 
mastectomy.
* Warning: Do not apply the cuff over a wound; otherwise it can cause further injury.
*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied 
around simultaneously, because this could cause temporary loss of function of those 
simultaneously-used monitoring ME equipment.
*On the rare occasion of a fault causing the cuff to remain fully inflated during 
measurement, open the cuff immediately. Prolonged high pressure (cuff pressure 

 

300mmHg or constant pressure 

 15mmHg for more than 3 minutes) applied to the arm 

may lead to an ecchymosis.
*Please check that operation of the device does not result in prolonged impairment of 
patient blood circulation.
* When measuring, please avoid compression or restriction of the connection tubing.

CAUTION

* The device cannot be used with HF surgical equipment at the same time.
* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER 
was clinically investigated according to the requirements of ISO 81060-2:2013.
* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact 
V.MAX distributer or LOCAL DEALER.
* This device is contraindicated for any female who may be suspected of, or is pregnant. 
Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
* Too frequent and consecutive measurements could cause disturbances in blood 
circulation and injuries.
* This unit is not suitable for continuous monitoring during medical emergencies or 
operations. Otherwise, the patient’s arm and fingers will become anesthetic, swollen and 
even purple due to a lack of blood.
* When not in use, store the device in a dry room and protect it against extreme moisture, 
heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
* This device may be used only for the purpose described in this booklet. The manufacturer 
cannot be held liable for damage caused by incorrect application.
*This device comprises sensitive components and must be treated with caution. Observe 
the storage and operating conditions described in this booklet.
* The equipment is not AP/APG equipment and not suitable for use in the presence of a 
flammable anesthetic mixture with air with oxygen or nitrous oxide.
* Warning: No servicing/maintenance while the ME equipment is in use.
* The patient is an intended operator.
* The patient can measure, change batteries under normal circumstances and maintain the 
device and its accessories according to the user manual.
* To avoid measurement errors, please avoid the condition of strong electromagnetic field 
radiated interference signal or electrical fast transient/burst signal.
*The blood pressure monitor, and the cuff are suitable for use within the patient 
environment. If you are allergic to polyester, nylon or plastic, please don't use this device. 
* During use, the patient will be in contact with the cuff. The materials of the cuff have been 
tested and found to comply with requirements of ISO 10993-5:2009 and ISO 
10993-10:2010. It will not cause any potential sensation or irritation reaction.
* Adaptor is specified as a part of ME EQUIPMENT.
* If you experience discomfort during a measurement, such as pain in the arm or other 
complaints, press the START/STOP button to release the air immediately from the cuff. 
Loosen the cuff and remove it from your arm.
* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. 
Should the cuff not deflate when pressure reaches 40 kPa (300 mmHg), detach the cuff 
from the arm and press the START/STOP button to stop inflation.
* Before use, make sure the device functions safely and is in proper working condition. 
Check the device, do not use the device if it is damaged in any way. The continuous use of 
a damaged unit may cause injury, improper results, or serious danger.
* Do not wash the cuff in a washing machine or dishwasher!
* The service life of the cuff may vary by the frequency of washing, skin condition, and 
storage state. The typical service life is 10000 times.

Summary of Contents for TMB-1776

Page 1: ...Pressure Monitor TMB 1776 Please read the user s manual carefully and thoroughly so as to ensure the safe usage of this product Keep the manual well stored for further reference in case you have probl...

Page 2: ...ds Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances Contraindications ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood p...

Page 3: ...igns below might be in the user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for COMPLIES WITH MDD 93 42 EEC REQ...

Page 4: ...ation When measuring please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disc...

Page 5: ...two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact your LOCAL DEALER for SERVICE Don t open or repair the device by yourself...

Page 6: ...he battery cover Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up The Choice of Power...

Page 7: ...re using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory The setting range of the year 2017 2057 time format available 24H 1 When the monit...

Page 8: ...rtably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep br...

Page 9: ...ll finish the whole measurement Take user 1 for example 2 Press the to power off otherwise it will turn off within 1 minute The current No is No 1 The corresponding time is 6 00 24H Format The corresp...

Page 10: ...or 3 seconds when the monitor is in the memory recall mode dEL ALL will flash on the display 2 Press to confirm deleting and dEL donE will be shown on the display then the monitor will turn off 3 If y...

Page 11: ...he device is not suitable for use on patients with implanted electrical devices such as cardiac pacemakers and defibrillators Ask your Doctor before use 3 We suggest above mentioned population to meas...

Page 12: ...arging Systolic relax blood entering Diastolic When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When...

Page 13: ...ement can affect the measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the...

Page 14: ...anual General Requirements for Safety Electromagnetic compatibility Performance requirements Clinical investigation Usability Software life cycle processes Bio compatibility ISO 10993 1 2009 Biologica...

Page 15: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Page 16: ...ion appropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output po...

Page 17: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Reviews: