QA Maintenance and Inspection ViZion DR+ Detector Guide v2.0
Opal
1
Customer Support
1.800.366.5343 –
VZDR-UG-QA Maintenance and Inspection ViZion DR+ Detector Guide-D-20141012-2.0
QA Maintenance and Inspection
Vizion DR+ Detector Guide
For safety reasons, be sure to turn OFF the power of each instrument when the performed inspection
indicates power to be off. Failing to power off instrument when requested may result in electric shock.
When the instrument is cleaned, be sure to turn OFF the power of each instrument, and unplug the
power supply cord from the AC outlet. Never use benzene, thinner or any other flammable cleaning
agents, as this may result in fire or electric shock.
The instrument must be repaired by a qualified engineer only. Failure to do so may result may result in
fire, electric shock, or accident.
I.
Environment of Use and Storage
Do not install the instrument in a location with the conditions listed below. Otherwise, it may result in
failure or malfunction, fall or cause fire or injury.
Close to facilities where water is used.
Where it will be exposed to direct sunlight.
Close to air-conditioner or ventilation equipment.
Close to heat source such as a heater.
Prone to vibration.
Insecure place.
Dusty environment.
Saline or sulfurous environment.
High temperature or humidity.
Do not place the storage case or instrument in a location with the conditions listed below.
Where the cable of the sensor unit will be strongly pulled when the sensor unit is put into the
case, otherwise, the cable may be damaged, resulting in fire or electric shock.
Where someone might get their foot caught in the cable of the sensor unit is put in the case.
Otherwise they could trip over, resulting in injury