ViVitro Labs SuperPump AR Series User Manual Download Page 7

 

 

Document No. 17476-C 

© ViVitro Labs Inc. 2018 

Page 7 of 22 

SuperPump AR System User Manual 

 

3  Installation 

The SuperPump System has several steps to the main mechanical sub-assemblies, and a few 
steps for the electrical components installation/configuration.  

NOTE:  It  is  critical  for  the  assembly  steps  to  be  followed  in  order  to  assure  successful 

operation. 

3.1 

Unpacking the SuperPump 

The  SuperPump  AR  arrives  in  2  containers;  a  wooden  shipping  crate  for  the  Pump  and  a 
cardboard  container  for  the  Controller.  Inspect  the  shipping  containers  for  external  damage. 
Retain the shipping carton and packing materials for possible re-shipment. Inspect the system 
components for  damage. If there is evidence of damage contact the carrier, initiate a damage 
claim, and inform ViVitro Labs. 

The SuperPump arrives pre-assembled with the following parts: 

  Mounting base 

  Linear Actuator with Cable 

  Actuator Extension Cable 

  SuperPump Controller 

3.2 

Installing the SuperPump  

The SuperPump is supplied mounted to a rigid plastic base for use on a bench top. Optionally 
the  Pump  may  be  mounted  directly  to  a  structure  or  test  apparatus  using  #10-32  fasteners 
through the mounting feet of the Pump. See bolt pattern layout in section 10. 

WARNING:  Any  accessory  that  is  attached  to  the  SuperPump  must  be  tested  to  375  mmHg 
pressure. 

The SuperPump must be mounted horizontally. The back side of the Pump cylinder is provided 
with  a  drain;  however  vertical  mounting  of  the  Pump  can  result  in  water  entering  the  ball 
screw/motor causing an electrical hazard. 

 

The  Controller  should  be  located  away  from  the  Pump  to  avoid  any  water  or  test  fluid.  An 
extension cable is provided to allow extra separation if required. 

It is recommended that the Pump act on the test fluid through a membrane to maintain isolation 
between  working  and  test  fluid.  The  piston  should  not  be  directly  exposed  to  blood  analog 
solution or saline. Distilled water is recommended as the working fluid in contact with the piston. 

When  connecting  the  Controller  to  other  equipment  such  as  computers  or  a  signal  generator 
ensure  all  equipment  is  on  the  same  ground.  Ground  loops  can  cause  noise  or  signal  offsets 
which  will  result  in  a  noisy  or  inconsistent  waveform.  Any  test  setup  should  be  validated  by 
running known samples before testing. 

 

 

 

Summary of Contents for SuperPump AR Series

Page 1: ...Document No 17476 C ViVitro Labs Inc 2018 Page 1 of 22 SuperPump System User Manual AR Series ...

Page 2: ...cal photocopying recording scanning or otherwise without the prior written permission of the Publisher Requests to the Publisher for permission should be addressed to ViVitro Labs Inc 455 Boleskine Rd Victoria BC V8Z 1E7 Canada tel 250 388 3531 fax 250 483 1975 e mail support vivitrolabs com Limit of Liability Disclaimer of Warranty While the publisher and author have used their best efforts in pr...

Page 3: ...ulatory compliance What we offer ViVitro Labs respected clients come from all over the world and include leaders in the cardiovascular implant industry Academic Research ViVitro Labs is a leading vendor to academic labs worldwide Hundreds of citations refer to the use of ViVitro Labs equipment For example the ViVitro Labs SuperPump has become a worldwide standard for the generation of physiologica...

Page 4: ...ring On the System 9 4 2 Front Panel 10 4 3 Rear Panel 10 4 4 Select Waveform 11 4 5 Sync Pulse Delay 11 4 6 Cycle Count Totalizer 12 4 7 Pump Operation 12 5 Waveforms 13 5 1 Internal Waveforms 13 5 2 External Waveforms 13 6 Faults and Alarms 14 6 1 Limit Switches 14 6 2 Pump Error 14 7 Maintenance and Cleaning 15 7 1 Cleaning 15 7 2 Piston Seal Leakage 15 7 3 Piston Seal Replacement 15 8 Specific...

Page 5: ...he risk of shock do not open any covers that are fastened with screws Do not allow fluids to enter the interior of the instrument In the event of such a spill disconnect the power cable before cleaning up Do not operate SuperPump in extreme humidity or in conditions that can create condensation Protect the instrument against dust and moisture and avoid physical shock and strong forces Any accessor...

Page 6: ...studies throughout the world Visit the Citation Database on vivitrolabs com for the current list of applications and articles The SuperPump provides maximum operator control with precision accuracy The Pump can be controlled by preconfigured waveforms or by customizable ViVitest software input 2 2 Features The SuperPump AR series from ViVitro Labs Inc features include Piston in cylinder Pump head ...

Page 7: ...tionally the Pump may be mounted directly to a structure or test apparatus using 10 32 fasteners through the mounting feet of the Pump See bolt pattern layout in section 10 WARNING Any accessory that is attached to the SuperPump must be tested to 375 mmHg pressure The SuperPump must be mounted horizontally The back side of the Pump cylinder is provided with a drain however vertical mounting of the...

Page 8: ... All equipment be checked for correct fuse ratings prior to use See section 3 4 3 4 Fuses When using the ViVitro Labs SuperPump with 220V 240V power the Controller fuses must be replaced with the 2A fuses provided If the Controller is being used with 100V 127V power the 4A fuses shipped in the device must be used Fuse Requirements 100V 127V Power 2x 4A Fuses included in Controller 220V 240V Power ...

Page 9: ... 0 5 2014 02 20 0 0 Gain 0 0 ml Stroke Initializing After automatically homing the Pump the screen will appear as below The most recently used waveform source will appear in the Current Waveform section If the Pump is not connected to the Controller when the system is powered on Pump Error will be displayed Power off the Controller connect the Pump cable and power the Controller on to clear the er...

Page 10: ... the connectors From left to right Waveform In BNC connector used when an external signal is used as the waveform Position BNC connector analog output signal for the piston position dl dt BNC connector analog output signal proportional to instantaneous flow rate Sync Out BNC connector 5V TTL signal triggered at the rising zero crossing of the waveform The timing of the signal can be adjusted to an...

Page 11: ...al clockwise and press the dial to select the displayed waveform Refer to section 4 7 for more details on waveform options 0 0 0 Gain 0 0 ml Stroke Wfm PHYSIO_70 4 5 Sync Pulse Delay The Sync Pulse is a 5V signal with 10ms duration which can be used for synchronizing and triggering equipment such as high speed cameras Press the Menu Selector Dial Turn the menu dial until 2 Sync Pulse Delay is on t...

Page 12: ...in 0 0 ml Stroke 3 Totalizer 4 7 Pump Operation 0 0 0 Gain 0 0 ml Stroke PHYSIO_70 0 msec Turn the Amplitude dial until the desired stroke volume is reached For example 51 40 7 Gain 73 4 ml Stroke PHYSIO_70 0 msec To pause or stop the Pump either press the Amplitude dial Once the Amplitude dial has been pressed once the system is paused and the following screen will be displayed 57 0 0 Gain 73 4 m...

Page 13: ...ernal see section 4 4 for the SuperPump to operate using the external input The external waveform must conform to the following specifications Amplitude 5 0 V maximum Any waveform that exceeds this amplitude will not be accurately represented in the Pump movement Rate Frequency 3 bpm beats per minute minimum to 200 bpm maximum 0 05 Hz to 3 33 Hz NOTE Do not input waveforms with frequency greater t...

Page 14: ...ain 180 0 ml Stroke Limit reached 6 2 Pump Error If the Controller loses communication with the Pump or if the Pump is unable to move a Pump error message will appear If this occurs 1 Turn off the system 2 Check that the cables are properly connected 3 Ensure that the Pump is free of mechanical interference and damage 4 Turn the Controller on and allow it to initialize normally 5 If the alarm pers...

Page 15: ...7 2 Piston Seal Leakage Periodically the glass cylinder should be inspected to check for signs of water leaking past the piston seal During the initial stages of piston leakage a slight wipe mark of fluid may be seen on the inside of the cylinder Excessive leakage may lead to fluid ingress into the Pump electronics causing damage to the system and electrical hazard See section 7 3 for directions o...

Page 16: ...on 08981 and end cap from the Pump assembly 4 Carefully remove the glass cylinder from the piston CAUTION DURING THIS STEP BE CAREFUL WHEN HANDLING THE GLASS CYLINDER AS THE EDGES MAY BE SHARP 5 Remove the M10 nut and lockwasher from the front of the piston 6 Remove the SuperPump Piston 09532 from the rod end 20 x 4 5 SHCS Short Extrusion M10 Nut and Lock Washer End Cap Piston Seal SPM043 1 Piston...

Page 17: ...ly by hand 9 Clean any dust and debris from the glass cylinder Apply two layers of electrical tape on either side of one end of the glass cylinder as shown below 10 Push the piston into the glass cylinder SPM042 using the two layers tape as a glide point Remove the electrical tape from the Glass Cylinder SPM042 11 Slide rod end into the SuperPump Piston 09532 Fasten in place with a M10 Lock Washer...

Page 18: ... of the glass with the aluminum parts Tighten the fasteners gradually to prevent uneven stress on the glass cylinder Compress the sides of the Short Extrusion 08981 while tightening so that the sides sit flush with the end cap faces 17 Check for damage on O Ring 09534 Press the o ring into the recess of the SuperPump Piston Cap 09533 if not already inserted 18 Fasten the SuperPump Piston Cap 09533...

Page 19: ...oke Inputs External Waveform 5 V 0 05 Hz to 3 33 Hz 3 to 200 beats per minute Outputs Position 5 to 5 V 200mV mm dl dt 5 to 5 V 539 ml s V Sync Pulse 5V signal for 10ms duration Adjustable from 0 to 2000ms delay LE at rising zero crossing Electrical Requirements Voltage 90 240 VAC 50 60Hz Power Max 300W Dimensions Controller 11 x 48 x 33 cm Pump 15 x 55 x 17 cm Environmental Operating Temperature ...

Page 20: ...or or actuator If the system is over pressurized the current output to the motor will be regulated resulting in the stroke volume being limited There are two ways to determine if the pressure limit is reached This will depend on whether the SuperPump has reached the instantaneous current limit or the continuous current limit If the instantaneous current limit is reached the Stroke Volume mL stroke...

Page 21: ...joints in tubing such as reducers tees and elbows If possible place them after the point of pressure measurement in the flow loop Ensure there is sufficient compliance and or a reservoir in the flow loop to accommodate the volume displacement of the SuperPump piston s stroke volume If it is suspected that the current pressure limit is being reached it is recommended to monitor the pressure being g...

Page 22: ...Document No 17476 C ViVitro Labs Inc 2018 Page 22 of 22 SuperPump AR System User Manual 10 Mounting Details 10 1 Mounting Feet Bolt Pattern 10 2 Pump Cylinder Mounting Face Bolt Pattern ...

Reviews: