Probable Cause

Problem

Recommended Action

Batteries are drained.

Replace all batteries with new ones .

Replace all batteries with new ones .

Replace all batteries with new ones .

Warm up the batteries, or use the

device in a warmer setting.

Adjust patient and wrist cuff

to measure.

Relax, and try to take a

measure again.

Rest a while, relax and

measure again.

Check that the wrist cuff is

fastened properly and retake

the measurement.

Refrain from moving

during measurement.

Do not take measurements after

exercise or coming back from

the outdoors.

Push the power button again,

and then begin measure again.

Use Alkaline batteries and replace

all batteries at the same time

with same brand batteries.

Reinsert batteries in the

correct position .

Batteries are drained.

Batteries are drained.

Perhaps the cuff was 

improperly positioned .

Blood pressure naturally varies

throughout the da y.

Bodily movement during device use.

The wrist cuff is not fastened properly .

It could be a normal action if the

user’s blood pressure is higher than

the initial pressure value, the device

automatically pumps to a higher

pressure by 40mmHg each time.

Measurement shortly after exercise

or exposure to the outdoors.

System design.

In colder temperatures batteries

have weaker electrical charges. 

Different battery brands have

different life spans.

Batteries are not correctly

aligned with terminals.

Low Battery

Symbol appears.

Nothing appears in

the display even

when the power is

turned on.

Suspicious heart

rate results,

Suspicious blood

pressure results,

No reading after

measurement.

Power switches off

automaticall y.

Device operation

time is inconsistent.

During measuring,

air reinflates.

© Copyright and Distributed by Vive Health
8955 Fontana Del Sol Way, Naples, FL 34109
All Rights Reserved. Made in China.
Email: [email protected]

ERROR CODES

TROUBLESHOOTING

FCC STATEMENT

This device complies with Part 15 of the FCC Rules. Operation is subject to the 
following two conditions: (1) this device may not cause harmful interference, and 
(2) this device must accept any interference received, including interference 
that may cause undesired operation. 

SPECIFICATIONS

•  2 AAA batteries required (Included)
•  Screen Size: 3cm x 3.8 cm (1.18” x 1.5”)
•  Weight: 120g (excluding battery)
•  Provides Systolic and Diastolic Pressure readings
•  Memory: Records 60 measurements 

•  Shows average measurement on bottom left

•  Adjustable Cuff range: 12.5cm - 21cm (4.9” - 8.3”)

•  Secure using velcro

•  Measurement Range:

•  Pressure: 30 - 280 mmHg Pulse
•  Pulse: 40 - 199 / minute

•  Calibration Accuracy:

•  Pressure: +-3mmgh 
•  Pulse: +- 4% of reading

WARNING

• 

This device is intended for adult indoor use only,

• 

This device is intended for non-invasive measuring and monitoring of arterial 
blood pressure. It is not intended for use on extremities other than the arm or 
for functions other than obtaining a blood pressure measurement.

• 

Do not confuse self-monitoring with self-diagnosis. This unit allows you to 
monitor your blood pressure. Do not begin or end medical treatment. 
Consult your physician for treatment or advice. 

• 

If you are taking medication, consult your physician to determine the 
most appropriate time to measure your blood pressure. Never change a 
prescribed medication without consulting your physician.

• 

If the cuff pressure exceeds 40kPa (300 mmHg), the unit will automatically 
deflate.  Should  the  cuff  not  deflate  when  pressure  exceeds  40  kPa  (300 
mmHg), detach the cuff from the arm and press the START/STOP button to 
stop inflation.

• 

To avoid measurement errors, carefully read this manual before using 
the product. 

• 

The equipment is not AP/APG equipment and not suitable for use in the 
presence  of  a  flammable  anesthetic  mixture  with  air  of  with  oxygen  or 
nitrous oxide.

• 

The operator shall not touch output of batteries and the patient simultaneously.

• 

To avoid measurement errors, please avoid the condition of strong 
electromagnetic field radiated interference signal or electrical fast transient/
burst signal when using the AC adapter.

• 

The user must check that the equipment functions safely and see that it is in 
proper working condition before being used.

• 

Please  use  ACCESSORIES  and  detachable  parts  specified/  authorized  by 
MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the 
user/ patients.

• 

Manufacturer will make available on request circuit diagrams, component 
parts list, etc.

• 

This unit is not suitable for continuous monitoring during medical emergencies 
or  operations.  Otherwise,  the  patient’s  arm  and  fingers  will  become 
anesthetic, swollen and even purple due to a lack of blood.

• 

Please use the device under the environment which was provided in the 
user manual. Otherwise, the performance and lifetime of the device will be 
impacted and reduced.

• 

During using, the patient will contact with the cuff. The materials of the cuff 
have been tested and found to comply with requirements of ISO 10993-
5:2009 and ISO 10993-10:2010. It will not cause any potential allergic 
reaction or contact injury.

• 

The device doesn’t need to be calibrated in two years of reliable service.

• 

Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT 
according to the local guidelines.

• 

When the device was used to measure patients who have common 
arrhythmias such as atrial or ventricular premature beats or arterial 
fibrillation, the best result may occur deviation. Please consult your physician 
about the result.

• 

The device is contraindicated for any female subject who may be

        suspected of, or is pregnant. Besides provided inaccurate readings, the
        effects of this device on the fetus are unknown.

• 

When using this device, please pay attention to the following situation which 
may interrupt blood flow and influence blood circulation of the patient, thus 
cause harmful injury to the patient: Too frequent and consecutive multiple 
measurements; The application of the cuff and its pressurization on any arm 
where intra-vascular access or therapy, or an arteriovenous (A-V) shunt, is 
present; Inflating the cuff on the arm on the side of a mastectomy.

• 

Do not apply the cuff over a wound, otherwise it can cause further injury.

• 

Don’t kink the connection tube, otherwise, the cuff pressure may continuously 
increase which can prevent blood flow and result in harmful injury to 
the PATIENT.

• 

The patient is an intended operator. The patient can measure under normal 
circumstances and maintain the device and its accessories according to the 
user manual.

• 

The blood pressure monitor, and the cuff are suitable for use within the 
patient environment. If you are allergic to dacron or plastic, please don’t use 
this device.

• 

Please keep the unit out of reach of infants, children or pets, since inhalation 
or swallowing of small parts is dangerous or even fatal.

• 

The device is not suitable for public use.

• 

The device is not intended for PATIENT transport outside a healthcare facility.

• 

The device cannot be used with HF surgical equipment at the same time. 
 

EMC GUIDANCE

1. MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and 
needs to be installed and put into service according to the EMC information 
provided in the ACCOMPANYING DOCUMENTS

2. Wireless communications equipment such as wireless home network devices, 
mobile phones, cordless telephones and their base stations, walkie-talkies can 
affect this equipment and should be kept at least a distance d=3, 3m away from 
the equipment.

(Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical 
cell phone with a maximum output power of 2 W yields  d=3, 3m at an IMMUNITY 
LEVEL of 3V/m)

GUARANTEE REGISTRATION

You are protected by Vive Health’s industry leading guarantees
and customer service:

If you did not purchase through vivehealth.com, please register at vivehealth.
com/register to validate your guarantee.

Product Code: DMD1029WHT

HIGH

NORMAL

110

1201

30 1401

60

180

Systolic Blood Pressure

Optimal

Normal

High-Normal

Mild Hypertension (Grade 1)

Moderate Hypertension (Grade 2)

Severe Hypertension (Grade 3)

Diastolic Blood Pressure

Hypertension Indica tor

mmHg

mmHg

100

90

85

80

HIGH

NORMAL

HIGH

NORMAL

HIGH

NORMAL

HIGH

NORMAL

HIGH

NORMAL

2 YEAR GUARANTEE

Err Cod

Meaning

No Pulse or detected pulse not strong.

Leakage in Cuff Pressure/Inflation too lo w.

Pressure fault.

Deflation fault.  

Memory Error.

Corrective Action

Err 0

Err 2

Err 3

Err

Err 1

Take off heavy clothes and try again.

The wrist cuff is not fastened properly . Re-apply
the cuff, and take measurement again.

Take off batteries to reboot the device, then
take another measurement .

Low Battery.

Replace all batteries with new ones.

Rest a while, relax and try again.
The wrist cuff is not fastened properly . Re-apply
the cuff and take measurement again.