background image

 

 

 

18 

 

(2-6-1)

 

VIEW TYPE: Select the frame type to show 
detected area.    Frame type: transparent square, 
grid, both open or close all.   

(2-6-2)

 

OSD VIEW: Select to show “MOTION DETECTED” 
when detect moving image. 

(2-6-3)

 

TIME: Set Frame pausing time when detect the 
image is not moving . 

(2-6-4)

 

RETURN: Return to last page or Save & End 
Menu. 

(2-7)

 

DEFAULT: Return to factory setting.   

(2-8)

 

RETURN: Return to last page or Save & End Menu. 

(3)

 

LANGUAGES: 16 languages are selectable.   

(4)

 

DEFFECT: Dead Pixel compensation.   

 

Summary of Contents for UM20B-CSZ064

Page 1: ...Version 1 0 UM20B CSZ064 User s Guide...

Page 2: ...emble microscope with stand 2 1 3 IR Remote controller 4 1 4 Microscope menu functions 9 1 5 Microscope introduction 23 1 6 Microscope Focus 26 1 7 Microscope accessory 28 2 Printed Notice 29 2 1 Main...

Page 3: ...ntrol 1 Power adaptor 1 Calibrator 1 S114 Stand Instruction Guide included 1 User s Guide 1 White Balance Card 1 1 Software download http www vitiny com vitiny_en ap_program html 1 2 Assemble microsco...

Page 4: ...3 1 2 2 Start to use microscope Connect the HDMI cable b and adaptor c to HDMI monitor and power See below a CSZ064 lens b c...

Page 5: ...onds 1 3 IR Remote controller The microscope operation and functions is controlled by the IR remote controller 1 Power Turn on off 2 HDMI When in HDMI mode the HDMI cable must connect with the screen...

Page 6: ...Zoom in Zooms in speed increases when held 7 2 Zoom out Zooms out speed increases when held 7 3 Step zoom in zoom in by steps 7 4 Step zoom out zoom out by steps 8 Autofocus Does not suggest to use t...

Page 7: ...e image is too dark 9 2 Decrease Use to decrease exposure when the image is too bright 10 Menu on Screen Open close menu for setting Refer to Section 1 4 11 Up Choose the previous item if Crosshair is...

Page 8: ...us and only show 3 seconds 16 1 Area e 16 2 1 OZ Optical Zoom Ex OZ 500 500 means sensor position 16 2 2 SZ Step Zoom Ex SZ 500 Same format as OZ 16 2 3 AFS Single Autofocus Ex AFS A500 A means search...

Page 9: ...Ex FRZ ON means freeze image FRZ OFF means unfreeze image 16 2 9 WB White Balance Calibration Ex WB SET means calibrate white balance 16 2 10 NONE No Function 17 Auto White Balance Calibration To cali...

Page 10: ...1 4 Microscope menu functions Press button to enter MENU press Up and Down to choose item Left and Right to turn on off or adjust parameter Enter means more functions press Menu to enter next function...

Page 11: ...10 1 4 1 EXPOSURE To set exposure time see below 1 Shutter Digitally control amount of light the amount is smaller the image is darker and the frame rate is also slower...

Page 12: ...ghtness 5 D WDR Digitally adjusting exposure in areas of the frame to maintain optimum detail in both the shadows and highlights of the image 5 1 LEVEL Adjust image exposure level from too dark or too...

Page 13: ...Return to Factory setting 6 4 RETURN Return to Last page or Save End Menu 7 RETURN Return to Last page or Save End Menu 1 4 2 BACKLIGHT Press left right button to choose turn on off Backlight mode 1...

Page 14: ...k Light Compensation To identify the over luminance area by backlight and proceed masking the area 3 1 SELECT Select the areas which needs to be adjusted 4 areas are selectable 3 2 DISPLAY Display the...

Page 15: ...urn to last page or Save End Menu 1 4 3 MANUAL WB Calibrate the white light balance to find out right White color 1 AWB Auto white balance 2 ATW Trace white area and auto calibrated 3 AWC SET Semi aut...

Page 16: ...ction NR To obtain high quality output image and increase file compression See below 1 2D NR LOW MIDDLE HIGH Edge Preserving 2D NR 2 3D NR LOW MIDDLE HIGH Motion Adaptive 3D NR 3 Return Return to last...

Page 17: ...right to turn on or off 1 1 FREEZE Turn ON OFF to freeze image If Freeze function is turn on the 1 2 Mirror function can not be used at the same time 1 2 MIRROR Image rotatory Horizontal or Vertical 1...

Page 18: ...nge as MIRROR function is ON 2 2 DISPLAY Display the selected area Position and area can be set in 2 1 2 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher s...

Page 19: ...Select to show MOTION DETECTED when detect moving image 2 6 3 TIME Set Frame pausing time when detect the image is not moving 2 6 4 RETURN Return to last page or Save End Menu 2 7 DEFAULT Return to fa...

Page 20: ...ill follow the threshold setting to adjust brightness and compensate the dead pixel 4 1 2 LEVEL Set the threshold value to judge if compensation is needed 4 1 3 RETURN Return to last page or Save End...

Page 21: ...udge if it needs to be corrected 4 2 5 AGC Auto Gain Control When light source is not enough AGC will follow the threshold setting to adjust brightness and compensate the dead pixel 4 2 6 SENS UP Rais...

Page 22: ...higher value increases the higher contrast along near edges of the image 1 1 LEVEL Adjust Image Sharpness level 1 2 START AGC Set start levels to adjust sharpness 1 3 END AGC Set end levels to end adj...

Page 23: ...l 2 4 RETURN Return to last page or Save End Menu 3 LSC Lens Shading Correction Use Lens shading correction to average the luminance of image 4 VIDEO OUT Select NTSC or PAL for TV output 5 RETURN Retu...

Page 24: ...n the IR receiver area 1 IR remote control aiming area f The indicator will blink whenever buttons are pressed If the indicator did not blink means the signal was not delivered and please repress agai...

Page 25: ...adaptor is only compatible with the cord in the box Please do not use other adaptor 2 USB j USB port for PC connection or Firmware update 3 HDMI K The HDMI port is used to connect the microscope to a...

Page 26: ...hen insert CSZ064 lens into Ring Adaptor l and screw tight with Hex Key Insert the lens with Ring adaptor into Universal Adaptor then screw tight 1 5 4 Illuminance LED Ring Light m Insert LED Ring Lig...

Page 27: ...to 4X Step 3 Adjust the knob o to find clear image and the working distance will be exactly 35mm to the object When the image is clear turn the optical zoom from 4x to 0 6x to see if image is clear at...

Page 28: ...27 n o...

Page 29: ...e will change 2x CSZ064 lens working distance 35mm 1 38 2x CSZ064 Lens on 24 monitor Reference data Optical Zoom 0 6x 1x 2x 3x 4x Fixed WD 35mm FOV mm 4 65 2 9 1 4 0 95 0 7 Mag X 101x 163x 380x 500x 6...

Page 30: ...lfunction if it encounters strong collision vibration or distortion 2 1 4 Turn the microscope off before cutting off the power supply do not forced cut off the power supply 2 1 5 Do not face the lens...

Page 31: ...3 0 Port Type C output YUV MJPE format Max res 1080P Focus Control IR remote control Power Supplier Adaptor Output DC 5 0V 2 0A Input AC 100 240V 50 60Hz A C Power Consumption 0 225A Max Size 280 L 66...

Page 32: ...e the device or the device is dropped or damaged please switch off and remove the power supply to avoid fire and electric shock Do not look steadily into Light sources after switching on the host as i...

Page 33: ...r etc Services caused by any of the following situations in the period of the warranty shall not be provided free of charge 1 Improper use or disassembly repair or refitting 2 Any damages to the devic...

Page 34: ...for Confirmation Stamp is necessary for validation of the Warranty Distributor s seal shall include name of the shop telephone and address Please ask the distributor to fill in the name of the shop ad...

Page 35: ......

Page 36: ...HTTP WWW VITINY COM MicroLinks Technology Corp All rights reserved...

Reviews: