background image

 

 

(16)

 

Status  Bar 

Suggest  to  OFF  the  Status  Bar  in  CSZ1236 

model. 

 

(17)

 

Auto  White  Balance  Calibration

:  To  calibrate  the  white 

balance, place the white balance card (included in box) under 
the lens, focus until clear, and choose On. Because the white 
balance is sometimes influenced by external light sources, it is 
recommended  to  calibrate  the  white  balance  if  the  color 
seems off or incorrect.

   

 

(18)

 

Image Freeze

: ON/Off to freeze or unfreeze the image.

 

Summary of Contents for UM08-CSZ1236

Page 1: ...Version 1 0A UM08 CSZ1236 User s Guide...

Page 2: ......

Page 3: ...ontents 2 1 2 CSZ1236 Lens Assembling 2 1 3 IR Remote controller 5 1 4 Microscope menu functions 8 1 5 Microscope introduction 21 1 6 Microscope Focus 25 1 7 Microscope accessory 28 2 Printed Notice 2...

Page 4: ...trol 1 Power adaptor 1 User s Guide 1 S107 Stand 1 CSZ1236 Lens 1 Stand assemble guide 1 Gooseneck Light 1 1 2 CSZ1236 Lens Assembling 1 2 1 Fix microscope Screw CSZ1236 Lens with UM08 Camera the adju...

Page 5: ...3 1 2 2 Start to use microscope Connect the HDMI cable c and adaptor d within the box to HDMI monitor and power See below a a b...

Page 6: ...4 1 2 3 Power On Off Press Power to turn on microscope...

Page 7: ...apshot No Function only with IMB05 5 LED Adjustment No Function 6 Motor Reset Recalibrates entire operation Returns to the lens position If a loss of electricity occurred during operation please reset...

Page 8: ...never image blurred 9 Exposure Value 18 levels of exposure value are selectable from 9 Auto 9 9 1 Increase Use to increase exposure when the image is too dark 9 2 Decrease Use to decrease exposure whe...

Page 9: ...e balance place the white balance card included in box under the lens focus until clear and choose On Because the white balance is sometimes influenced by external light sources it is recommended to c...

Page 10: ...There are 8 functions in menu press Up and Down to choose item Left and Right to turn on off or adjust parameter Enter means more functions press Menu to enter next function or confirm means no more...

Page 11: ...s smaller the image is darker and the frame rate is also slower 2 AGC Auto Gain Control auto control the brightness based on setting when light is not enough 3 SENS UP Only available when item 2 AGC v...

Page 12: ...n to see clear image Below fig for De Fog setting 6 1 POS SIZE Adjust de fog position and area 6 2 GRADATION To set gradation levels near the edge of defog area 6 3 DEFAULT Return to default setting 6...

Page 13: ...OFFSET Wide Dynamic Range OFFSET 2 3 RETURN Return to last page 3 BLC Back Light Compensation To adjust brightness for darker image in back light environment 3 1 GAIN To adjust brightness level 3 2 A...

Page 14: ...ng the area 4 1 SELECT Select the areas which needs to be adjusted 4 areas are selectable 4 2 DISPLAY Display the selected area 4 3 LEVEL Set value to judge if the image is too bright 4 4 MODE Super b...

Page 15: ...hite balance 2 ATW Auto tracing white area and calibration 3 AWC SET Semi auto white balance Use to track the white color from current image 4 MANUAL Manually adjust blue and red values to find correc...

Page 16: ...D NR 3 1 1 SMART NR The sensitivity to detect the moving object The lower value gets higher sensitivity 3 1 2 RETURN Return to last page 3 2 LEVEL The threshold to judge noise 3 3 START AGC Start auto...

Page 17: ...turn Return to last page 1 4 5 SPECIAL Set special functions 1 D EFFECT There are few more functions in D EFFECT but 1 3 1 4 and 1 6 function can only turn on one item at one time Press Left or right...

Page 18: ...chose area Digital zoom selection can from 1 62 times See below setting for D Zoom 1 3 1 PIP No function 1 3 2 D Zoom Set digital zoom times 1 3 3 PAN TILT To set the enlarge position 1 3 4 DEFAULT R...

Page 19: ...y enlargement frame Position and area can be set in 1 4 1 1 4 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher sensitivity 1 4 4 PIP No Function 1 4 5 D Zo...

Page 20: ...Select the areas which needs to be detected 4 areas are selectable 2 2 DISPLAY Display the selected area Position and area can be set in 2 1 2 3 SENSITIVITY The sensitivity to detect whenever object m...

Page 21: ...Pixel Correction Correct the live dead pixel during preview 4 2 1 LEVEL Set the threshold value to judge if it needs RETURN Return to last page 4 2 STATIC DPC Static Dead Pixel Correction scan and fi...

Page 22: ...the dead pixel 4 2 4 AGC Auto Gain Control When light is not enough AGC will follow the threshold setting to compensate the dead pixel 4 2 5 RETURN Return to last page 4 2 2 to be corrected 5 VERSION...

Page 23: ...ctory default setting Press Left right button to choose CHANGE to return default 1 4 8 RETURN Exit from menu 1 5 Microscope introduction 1 5 1 IR Remote Area Indicator When both f and g LED indicator...

Page 24: ...blink whenever buttons are pressed If the indicator did not blink means the signal was not delivered and please repress again 2 HDMI indicator f When lit up the microscope is in HDMI mode If light bli...

Page 25: ...Connect the microscope to the screen using the HDMI cable from the box For TV screen connections make sure to choose the homologous image ratio and format 2 USB i To use with accessory IMB04 for focu...

Page 26: ...24 1 5 4 Illuminance Gooseneck Light l Insert the Gooseneck light clip into base near the pole tighten the screws Connect power with UM08 s power adaptor k...

Page 27: ...the contrary the smaller magnification the wider FOV and longer working distance There are 3 adjustment screws on CSZ1236 Lens The function as following A Focus Adjustment Loose the screws and rotate...

Page 28: ...ller aperture C Focal Length Adjustment The focal length is from12 36 each adjustment requires to adjust A Focus If the image is not clear after adjusted A and C means the focus distance is not correc...

Page 29: ...distance 300mm 20mm 24 monitor Ref data WD mm 300mm 200mm Aperture 4 4 Focal length 12 15 20 26 36 12 15 20 26 36 FOV mm 141 5 115 89 70 8 53 5 98 80 63 50 5 39 Mag X 3 4 4 1 5 3 6 7 8 9 4 8 5 9 7 5...

Page 30: ...1 7 Microscope accessory 1 7 1 White balance card The white side is used to calibrate the white balance If the object is tiny place it onto the white balance card and move the card instead of the obje...

Page 31: ...lfunction if it encounters strong collision vibration or distortion 2 1 4 Turn the microscope off before cutting off the power supply do not forced cut off the power supply 2 1 5 Do not face the lens...

Page 32: ...or DC Input HDMI port Output 1080P 1920 1080 Pixel image Focus control IR remote control Power supplier Adaptor Output DC 5 0V 2 0A Input AC 100 240V 50 60Hz A C Power Consumption 0 225A Max Size 115...

Page 33: ...the device or the device is dropped or damaged please switch off and remove the power supply to avoid fire and electric shock Do not look steadily into Light sources after switching on the host as it...

Page 34: ...r etc Services caused by any of the following situations in the period of the warranty shall not be provided free of charge 1 Improper use or disassembly repair or refitting 2 Any damages to the devic...

Page 35: ...for Confirmation Stamp is necessary for validation of the Warranty Distributor s seal shall include name of the shop telephone and address Please ask the distributor to fill in the name of the shop ad...

Page 36: ...HTTP WWW VITINY COM MicroLinks Technology Corp All rights reserved...

Reviews: