VistaLab ali-Q 2000-0005 Operating Manual Download Page 3

P a g e

 | 

3

 

Safety Instructions 

Read this operating manual completely before using the device for the first time. Also, observe the instructions for use of the 
accessories. This instrument may sometimes be used with hazardous materials, operations, and equipment. It is beyond the scope 
of this manual to address all the potential safety risks associated with its use in such applications. It is the responsibility of the 
user of this instrument to consult and establish appropriate safety and health practices and determine the applicability of 
regulatory limitations prior to use

ali-Q

 

is an aliquoting pipet controller designed for aspirating and dispensing aqueous solutions with plastic or glass serological 

pipets in 

1

 to 100 mL volumes. It is intended for laboratory use. Any use of this instrument in a medical or IVD setting is the sole 

responsibility 

of

 the user. 

! Please read the following carefully!

1.

Every user must read and understand this operating manual before operation.

2.

Follow general instructions for hazard prevention and safety instructions; e.g., wear protective clothing, eye protection and

gloves.

3.

Observe all specifications provided by reagent manufacturers.

4.

Never use the instrument in an atmosphere with a danger of explosion. Highly flammable liquids must not be pipetted.

5.

Only

 

use the instrument for pipetting liquids, with strict regard to the defined limitations of use. If in doubt, contact VistaLab.

6.

Always use the instrument in such a way that neither the user nor any other person is endangered. Avoid splashes. Only use

suitable vessels.

7.

Never use force on the instrument.

8.

Use only VistaLab accessories and spare parts. See Ordering Information on pg 18. Do not attempt to make any technical

alterations. Do not dismantle the instrument any further than is described in the operating manual! 

9.

Always check the instrument for visible damage before use. If there is a sign of a potential malfunction, immediately stop

pipetting. Consult the 'Troubleshooting' section of this manual or contact VistaLab if needed.

10.

Only the original Power supply can be used for recharging of the Lithium Ion battery. Do not replace with non-rechargeable 

batteries or rechargeable batteries of other manufacturers.

11.

The Power supply must be protected against moisture and must be used only for this instrument.

12.

Only authorized service personnel may service the instrument. Unauthorized repair or service may void the units’ warranty

13.

DO NOT AUTOCLAVE ali-Q. Only approved parts and accessories may be removed from unit for autoclaving. See pg. 12,

Cleaning / Decontaminating.

14.

WARNING! Improper use of the instrument or the battery (short-circuit, mechanical damage, overheating etc.) may cause

the

 

explosion of the battery.

Operating Conditions: 

o

Operating Temp: 15°C to 40°C (59°F to 104°F)

o

Storage Temp: -20°C to + 50°C (-4°F to +122°F)

o

Altitude up to 2000 m

o

Relative Humidity: 10% – 95%, non-condensing

o

Atmospheric Pressure: 70 – 106 kPa (525 – 795 mmHg) 

o

Vapor Pressure up to max. 500 mbar. Aspirate slowly above 300 mbar, to prevent liquid boiling

o

Density up to 9.0 g/cm

3

ali-Q Body Chemical Compatibility for cleaning and decontaminating: 

Water, diluted ethanol or isopropanol. For more information on chemical compatibility, see the support area of our website, 
www.vistalab.com/support 

Safety Compliance:  

See Declaration of Conformity at end of manual 
EN 61010-1 General Safety Requirements 

CSA C22.2, No. 61010.1 

EN 61326 EMC Requirements  

UL 31111-1 

<- 

Back to T

.

O

.

C

.

Summary of Contents for ali-Q 2000-0005

Page 1: ...Operating Manual ali Q Aliquoting Pipet Controller Operating Manual or click here or scan QR code...

Page 2: ...ode 7 Aliquot Mode 8 Intelligent Measuring System 9 LED Indicators 9 Technical Specifications 10 Troubleshooting 11 Maintenance 12 Cleaning Decontaminating 12 Autoclaving 12 Replacing Membrane Filter...

Page 3: ...sories and spare parts See Ordering Information on pg 18 Do not attempt to make any technical alterations Do not dismantle the instrument any further than is described in the operating manual 9 Always...

Page 4: ...harging port Micro USB N A 4 Aliquot button Nylon 5 Aspirate button Nylon 6 Dispense button Nylon 7 Battery Lithium ion Lithium Polymer 8 Battery compartment cover PC ABS 9 Nozzle Cover Nylon 10 Nozzl...

Page 5: ...on verb To separate a volume of solution or suspension into aliquots Aspirate to fill with liquid Dispense to eject liquid Serological Pipet pipet disposable glass or plastic tube with volume graduati...

Page 6: ...d control In Conventional Pipetting Mode see next page the speed that the ali Q aspirates and dispenses liquid is controlled by the touch sensitive aspirate and dispense buttons The amount of force us...

Page 7: ...the bottom gray dispense button until bottom of meniscus is at the desired volume graduation ATTENTION Do not fill the pipet over the max volume Severe damage can occur if any liquid enters ali Q Dec...

Page 8: ...d 4 Aliquot Carefully position the tip of the full pipet over the first receiving vessel Press and hold the purple aliquot button and the LED turns solid GREEN Hold button until green LED turns off an...

Page 9: ...tions Temperature humidity and pressure all have significant effects on the performance of any liquid handling device EXCEPT for the ali Q ali Q can sense changes in ambient conditions and compensate...

Page 10: ...ANY brand serological pipet Optimal performance when using VistaLab Wobble not serological pipets Not for use with Pasteur pipets Mains power supply Input Voltage 100 240 VAC Frequency 50 60 Hz Input...

Page 11: ...Wobble not pipets will significantly decrease damage to the nozzle insert Out of Calibration Using different equipment different density liquid varying environmental conditions Membrane filter clogge...

Page 12: ...ation lamps NOTE The effects of the UV exposure will likely cause discoloration over time Autoclaving NEVER AUTOCLAVE ALI Q Only the nozzle cover nozzle insert and membrane filter can be detached from...

Page 13: ...cord into the charging port of the ali Q under the aliquot volume dial ali Q can s ll be used when plugged into charger Replacing the Ba ery Removing Battery Hard reset 1 Remove battery cover On the...

Page 14: ...n at least once a year or whenever conditions and equipment change In lab Verification Adjustment Users can verify the accuracy of the dispensed aliquots in four ways 1 Gravimetric Use a digital scale...

Page 15: ...used For warranty claims please contact Technical Support to receive Return Authorization Unit will be inspected for defects and repaired replaced as necessary at the sole discretion of VistaLab Cust...

Page 16: ...zle Cover 1 pack 9020 9005 Replacement Power Supply w Universal outlet adapters 1 pack 9020 9008 Wall Mount 1 pack Wobble not Serological Pipets Catalog Number Description Quantity 4090 0005 Wobble no...

Page 17: ...ervice life According to the Directive 2002 96 EC of the European parliament and of the Council on Waste Electrical Equipment WEEE of 27 January 2003 electronic equipment requires disposal according t...

Page 18: ...General Requirements UL 61010 1 Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including...

Reviews: