WARNING
• The user should test the product before use to ensure
proper functionality.
• The user should test the VALIANT™’s optical power using a
radiometer prior to use.
• An irradiance value 800-1750mW/cm2 is acceptable
for standard and ramp modes, and a value
1350-3300mW/cm2 is acceptable for boost mode
• The irradiance value should be measured at 10s within
a 20s cycle for standard and ramp modes.
• As with any heavily used medical device, the user needs to
ensure a functional backup is readily available.
•
DO NOT
look directly into the light output. The light
attenuating shield should always be used. Additional eye
protection is suggested.
•
DO NOT
expose soft oral tissues at close proximity.
Maintain a safe distance between the lens and the soft tissue.
• If using the VALIANT™ curing light in the Standard Mode and in
close proximity of the gingival tissue,
DO NOT
expose tissue
for more than 20 seconds.
• In Boost Power Mode,
DO NOT
expose soft oral tissue for more
than 9 seconds (3 cycles)
•
DO NOT
insert fingers, instruments, or other objects into the
rear of the handpiece when the battery pack is removed
•
DO NOT
autoclave the handpiece, battery pack, charging base,
or light attenuating shield
14