visilab MK30 User Manual Download Page 59

Appendix  4.  Troubleshooting  the  Meter

In cases where you meet troubles in getting the meter working correctly you may need assistance. The manufac-

turer is able to help you but to alleviate our burden, please check the following points before calling.

A. Symptom: No moisture signal to the PC, no contact with the MK30 program (or AK30Mini, IRMA7Mini

etc.), but the meter is working itself. Bluetooth operation only

1. Are you sure that the COMnn port you are using in the Bluetooth driver program is the same as the one the

PC program is using? 

This is absolutely the most frequent reason for communication problems with the meter

when using the Bluetooth option.

 Change the port selection on the program's Configuration page.

2. If done all right thus far and still no connection is established with the meter, press the reset button on the

Bluetooth master module connected to the USB and then go On-line in the program.

C. Symptom:

The meter is dead.

1. Is the meter correctly powered? Are you sure? Are the displays showing something? If yes, the meter is

receiving power and/or its batteries are all right but may have a weak charge. If not, check the power cable from

the battery charger unit. It should be connected to a 230VAC or 110VAC wall inlet. Let's suppose the meter

is not in Low Power mode. If not certain, press the Low Power key to force some effect. Do you hear any beep

when a key is pressed? If not, the meter is either off or the batteries are dead. If the batteries are empty, charge

the meter overnight and try again.

2. Is the cable with the connector placed into the socket in the meter's handle? If there is still no power to the

meter, the power source may be faulty (rare). Let a technician check it or contact the manufacturer. A voltage

close to +9 ...+24  Volts should be available in it at the connector pins. All fuses in  the system are self-

recovering and need no service.

3. Try to charge the battery for at least 5 hours with the meter turned off. If that does not help, there may be

some internal malfunctioning or damage. In some cases, the internal fuse has overheated and prevents both

battery charging and meter operation due to overload. Turn off the meter and try to charge it for at least 5

hours.  If that does not help, the battery is really dead and does not take charge anymore. It must be replaced.

Usually a dead battery does not prevent using the meter but does not keep its charge.

C. Symptom:

The moisture signal is incorrect.

1. Are you using the proper calibration table for it?

2. Have you made adjustments to the table used (with Adjust)? Could that explain this? Check the amount of

adjustment in the menu system, Calibration and Adjust.

3. Is the meter's position different from normal assembly? The sample could be seen differently if it is curved or

in a quite different angle or distance compared to the original position. Then, do the Adjust to correct it.

4. Have you accidentally changed the table's state from MULTI to SCALE? This is likely the most frequent

reason for an incorrect reading.

5. Is the head temperature too high? It should normally be below +45C. Has the meter been subjected to very

high ambient temperatures? If the temperature has risen too much, irreversible damage (or at least changes to

some important parameters) have been caused to the meter. In mild cases there is no serious damage but the

meter has gained some extra offset to its moisture signals.

If all attempts show no proper indication of a working meter, contact your dealer. Do not send back the

meter without permission.

Summary of Contents for MK30

Page 1: ...ETER USER S MANUAL Visilab Signal Technologies Oy Address Tel Sep ntie 4 358 45 635 4885 www visilab fi email info visilab fi FI 07230 Monninkyl Y 0631208 0 ALV Rek FINLAND VAT FI 06312080 M nts l Reg...

Page 2: ......

Page 3: ......

Page 4: ......

Page 5: ...Model MK30 Compact...

Page 6: ...Calibration when Working with a Climatic Chamber and a Balance 25 Adjusting a Table 26 The Procedure for Adjusting 26 Standardization Menu 27 Alarm Menu 28 Service Menu 29 Unit and Communications Men...

Page 7: ...Definition of the Data Transmission Protocol 53 Appendix 3 Technical Specifications of the Moisture Meter MK30 Regular and Compact applies mostly to MK30LP too 56 Appendix 4 Troubleshooting the Meter...

Page 8: ......

Page 9: ......

Page 10: ......

Page 11: ...Correct Incorrect Incorrect Incorrect Correct Compress Correct Compress loose material...

Page 12: ...Correct Incorrect Center the cylinder one sheet tension at edges Min 50 mm Incorrect Correct...

Page 13: ...measuring thick material any background any background Min 30 50 mm Incorrect Correct...

Page 14: ...one sheet Incorrect Incorrect Incorrect Correct...

Page 15: ...0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000 0 20 40 60 80 100 MK30 MK30 W BW g m2 moisture total...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ...les Auto ON Auto OFF Batch Ave Marker Range Flash Cal Up Down Recipe edit Filter Select Recipe Recipe TABLES 6 7 43 PAPER LOGOHOLD AUTO TABLE 0 TIME REEL 12 7 8 9 ESC 4 5 6 Menu 1 2 0 3 Enter Save Low...

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ...0 10 20 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 SCALE the default raw signal raw signal pulp plate 1000 g m2 moisture...

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Page 33: ......

Page 34: ...7 8 9 ESC 4 5 6 Menu 1 2 0 3 Enter Save Low Pwr Stat s BIGNUM Graph Select Select Charge Banks Tables Auto ON Auto OFF Batch Ave Marker Range Flash Cal Up Down Recipe edit Filter Select Recipe Recipe...

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ......

Page 40: ......

Page 41: ......

Page 42: ......

Page 43: ......

Page 44: ......

Page 45: ......

Page 46: ......

Page 47: ......

Page 48: ......

Page 49: ......

Page 50: ......

Page 51: ......

Page 52: ......

Page 53: ......

Page 54: ......

Page 55: ......

Page 56: ......

Page 57: ......

Page 58: ...zing some of its internal parameters The version Compact has somewhat narrower moisture range 0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000 0 20 40 60 80 100 MK30 MK30 W BW g m2 moisture t...

Page 59: ...e may be faulty rare Let a technician check it or contact the manufacturer A voltage close to 9 24 Volts should be available in it at the connector pins All fuses in the system are self recovering and...

Page 60: ......

Page 61: ...4 Adjust 6 Stdze At some occasions a security question is involved SURE 1 If you respond by pressing key 1 the action is continued else not In several submenus pressing the Save key will either save t...

Page 62: ...ampling interval Further submenus 32 Edit batch size Series submenus C Erase data in Banks 5 1 FLASH 2 ZERO 1 2 Flash calibra tion Automatic zeroing of moisture read ing Flash calibration submenu T 6...

Page 63: ...raw signal and inquire for the corresponding moisture 2 Step 1 3 Execute 1 0406RAW 1 2 3 2 Edit number of points in use in the table MULTI 1 Calibrate 2 Edit 3 Points 2 2 STEP 1 4 Signal 5 Moisture F...

Page 64: ...dze 4 Adjust 1 Set delta 0 0 19 AC YE 1 Edit deltavalue 0 0 Affects one table only 6 Standardize Drift 1 Offset 0 0 1 Edit drift offset 0 0 Affects all tables Calibration submenus Arrow keys UP DOWN a...

Page 65: ...LU 21 BASE60 Select from a list use arrow keys up down and press Enter SURE 1 Clear a table 2 Edit 1 Name 2 Libname Edit the calibra tion table name or the library name 3 19 AC YE Edit calibration tab...

Page 66: ...vel 2 3 Turn alarm feature on off Service 0 Sleep 8 Thead 9 More Check 5 Check head temperature 8 Turn sleep feature on off DIM 0 Alarm submenus Service submenus C C More 1 SaveCF 5 ErData 7 BAUD C 9...

Page 67: ...nd Bluetooth PIN Switch between unit options 2 1 Editunit numeric phone keys arrows 3 Initialize the Bluetooth 4 Edit address 1 default 1 1 Range of Curve 1 Low 2 High 3 Automatic Edit Low end 0 0 1 E...

Page 68: ...intensi ties and brightness for each display TABLE display 5 Select 123 RGB Bright Select Adjust RBG intensi ties and brightness for each display LowPower display 4 Select 123 RGB Bright Select Adjust...

Page 69: ...ent 9 Battery 57 C Calibration 56 Calibration and Material Libraries 17 Calibration menu 21 Connecting the Cables and Starting the Meter 8 Contents of the delivery bundle 39 CRCCCITT 53 D Data Flow in...

Page 70: ...nce 56 Measuring Speed 56 measuring state 20 Memory Banks 20 Menus and Settings 19 Moisture range 56 MULTI 20 21 MULTI Calibration 23 N Noise level 56 Notices for Measuring 36 O Opening the Meter 51 O...

Page 71: ...mode 17 sleeping feature 29 sleeping mode 57 Spacers 56 Special Features of Different Models 36 Stability 56 Standardization Menu 27 T TABLES 16 Technical Specifications of the Moisture Meter 56 text...

Page 72: ......

Reviews: