background image

TECHNICAL NOTES FOR USERS - NOTE #033

The Procedure for Calibration For Infrared Moisture Meters

page 3

Visilab Signal Technologies Oy

Sepäntie 4

Tel. +358-45-635 4885

FI-07230 Monninkylä

[email protected]

Y0631208-0 ALV Rek.

FINLAND

www.visilab.fi

VAT  FI06312080

Mäntsälä Reg. 365.258

true moisture and you can compare the two readings. You can then keep the sample under the meter and do the

adjustment to fine tune to force the meter reading exactly the same. The fine tuning (Adjust) is in AK30/40 in

option "9=More.." and "4=Adjust". Type in some small decimal number to see what happens. Iterate to get

the correct reading. Before leaving the menus, save the table and possibly also the new configuration the main menu,

if necessary.

Do not forget to download the library to your PC to save your valuable work to a safe folder.  Always modify the library name too.

If you will need to use the dry weight instead of the default total weight moisture, you would need to do the conversion for each

moisture reading according to:

   total-% * 100

   total-% * 100

   total-% * 100

   total-% * 100

   total-% * 100

dry-% = --------------------

dry-% = --------------------

dry-% = --------------------

dry-% = --------------------

dry-% = --------------------

                                       (100 - total-%)

                                       (100 - total-%)

                                       (100 - total-%)

                                       (100 - total-%)

                                       (100 - total-%)

Tips and Tricks

Tips and Tricks

Tips and Tricks

Tips and Tricks

Tips and Tricks

With thin papers (BW < 150 g/m2) the paper is usually behaving with one-sided infrared moisture meters so that the total moisture

can be measured and sidedness is not so important. For best results, measure both sides of the sample. However, if you have climatized

the sample properly, this does not give any extra information. This is sheet calibration and it is performed for highest accuracy over

a black aluminum plate, one sheet at a time. Never keep other papers under the sample with thin papers while measuring.

If the sides are different, on the other hand, like one side coated and the other not, then it is advised to keep the side information

apart from each other. They really behave slightly differently. One has to decide which side is important for practical measurements

and perform the calibration for that side alone.

With thin papers sometimes one would need instead of sheet calibration also the reel calibration. This has an infinite BW and it can

be simulated at calibration time easily. While measuring in the chamber, always keep a bunch of similar or identical papers as the

background over which the actual sample is placed. Keep the resulting bunch rather tight when doing measurements to avoid any

air gaps between them. The signal readings will now be those corresponding to reel measurements. In this way, you might be able

to generate two calibrations with the same effort, in simple steps: one for sheet (black background) and another for reel (stack

background).

With thick papers it does not matter if you have any other papers under the sample or not. If it does, then you should have both

sheet and reel calibrations.

While waiting the samples to climatize for hours, set the meter to LowPower mode to avoid generating extra heat to the chamber.

That would unnecessarily load the control system and possibly slow down the whole process.

Reviews: