Maintenance
6
2.3 Cleaning and Replacing UV Sensor
Mineral deposits and sediment may accumulate on the sensor window decreasing the UV energy detected. Good
maintenance of pre-treatment equipment will reduce the accumulation of residues. If the system indicates that the UV
intensity is low, one cause may be a stained quartz sleeve and/or sensor window.
Prerequisites:
•
Disconnect the main power source and allow the system to cool for 10 minutes.
•
Shut off water supply and drain all lines.
•
Depressurize the unit. Place a small pail near the openings to catch any spills.
•
Remove the UV lamp. Refer to
•
Remove and clean the quartz sleeve. Refer to
Note:
The quartz sleeve and UV sensor should be cleaned at the same time.
Procedure:.
WA R N I N G
The UV sensor is very delicate instrument. Extreme care is required when handling and cleaning. The sensor window itself is constructed from quartz
which is extremely fragile, be careful you do not chip or break this quartz window. Manufacturer’s warranty does not cover damage due to neglect or
misuse.
1
•
Disconnect the UV sensor
from the controller by
disconnecting the sensor
cable.
2
•
Remove the UV sensor and O-
ring by grasping the body of
the sensor and rotating it
counter-clockwise.
Commercial
Scale
Remover
3
0
30
15
45
30 mins
•
Submerge only the end of the
sensor in the commercial
scale remover for 30 minutes.
4
1
2
•
Clean the sensor with a cotton
swab and spray with water.
•
Re-insert the UV sensor along
with the O-ring until a water-
tight seal is achieved.