Part number:
OXY-2-DUR-IFU-1 Rev A
CONFIDENTIAL
Page 12 of 26
5.1.5
Step 5: Installation into Sheath
8.
The non-sterile operator drops the combined reusable main unit and battery pack into
the sterile sheath at the edge of the sterile field.
5.1.6
Step 6: Closing the Sheath
9.
The sterile operator closes the lid of the sheath until it latches.
At this point, the non-sterile operator has completed their part of the installation process.
Once the sheath is latched closed, it should not be opened again for the duration of the
procedure.
After installation, the system will verify the sheath installation process as it goes through the
below screens.
The device warm-up may take up to ten minutes. Keep the device away from heat sources (such
as bright lights) and from cold circulating air during the warm-up.
If the error ‘Missing Calibration’ appears on
-screen during device warm-up, the device cannot be
used. Please use a different reusable main unit for this procedure and contact ViOptix.