In order to guarantee a careful hygiene and cleaning it is advisable to
follow scrupulously the procedures hereunder reported:
WARNING:
Disconnect from the mains before cleaning the unit.
•
Do not allow liquids and water to enter inside the unit, to prevent
from shortcircuits and corrosion of inner parts.
•
Clean the painted surfaces, the accessories and the connection
cables using only a damp cloth and neutral detergents. Take care to
dry the parts with a dry cloth. Do not use corrosive or abrasive
solvents (alcohol, gasoline, "trielina").
•
The bite rod, the centering bite and the ear centering pins must be
replaced after each examination where they have been used. In
addition, the bite must be disinfected by using a 2% Glutharaldeid
solution, following the rules provided by the manufactures of the
cleaning solution itself.
•
The chin support, the patient positioning handles and the
temple-clamp must be carefully cleaned and disinfected with a 2%
Glutharaldeid solution or similar (find out what is available; e.g.
Milton) after each examination where they have been used.