4.4 Connection of the Sensor
For normal operating the sensor is connected via a M12 sensor connector (3 poles).
A suitable cable can be delivered optionally (lengths: 2 m/5 m/10 m). A shielding cable
is not necessary for short distances. We recommend a shielding for longer length
(> 30 m) or in rough industrial environments.
For the service a special cable (5 poles) is required. Using the additional lines it is
possible to configure the sensor and to adapt it to customized specifications. Do not
connect unused lines!
Use the service lines only in combination with the ABD Monitor for the
Air Bubble Detektor (delivered on request). Otherwise, the sensor could be
damaged.
Pin
Color
Connection
1
brown
Power supply v12 to 30 VDC
2
white
For service only: serial control input (5V/TTL)
3
blue
Ground
4
black
Switching output, PNP, max. 150 mA
5
grey
For service only: serial control output (5V/TTL)
shield
If available: should be grounded on the side of controlling unit
View on connector of the sensor
View on connector of the cable (female)
Pin configuration of M12 sensor connector
4.5 Settings/ABD Monitor
A computer in connection with the ABD monitor (consisting of USB data converter, M12
sensor cable, USB cable, CD with software ABD monitor, operating instructions ABD
monitor) is required for settings or for diagnostic purposes.
The sensor can be customised via parameters:
• Bubble sensitivity
• Reaction time/ hold time of the output/ output states
• internal regulation, e.g. to optimise the behaviour in case of pressure fluctuations
in the hose