Press

F

key to set carrier frequency, input

6

,

3

,

0

,

KHz/mV

in turn to set output frequency

Press

f

1

key to set the modulation signal as 1000Hz

Press

%

key and input

3

,

0

,

%/S

to set modulation depth

Press

V

key to set output amplitude, press

2

,

MHz/V

in turn to set output amplitude, and it completes.

During operation, the sequence is random to set carrier frequency, amplitude, depth and modulation signal.

4-2-3-3.

4-2-3-3.

4-2-3-3.

4-2-3-3.

OUTER-MODULATION

OUTER-MODULATION

OUTER-MODULATION

OUTER-MODULATION

SETTING

SETTING

SETTING

SETTING

When the inner-modulation parameter can not meet the requirement, you can set the instrument to outer-modulation mode. P ress

MOD

key to select outer-modulation mode, and

AM

and

EXT/LOG

indicator light, the instrument is in outer-modulation mode. Input the outer-modulation signal from

EXT

terminal

Press

F

key can set carrier wave

frequency

Press

V

key can set the amplitude of the output signal

Due to the modulation depth relates to the magnitude of the outer-modulation signal, the inner- modulation depth

setting will not take function. The basic setting is the same as that of inner-modulation, and it is unnecessary to set modulation depth and frequency of inner-modulation signal.

The seemly frequency of outer-modulation signal is between 100Hz

——

100KHz

the input impedance for

EXT

terminal is 600

Ω,

when input signal is 0dB, the modulation

depth is 50%; when input signal is +6dB, the modulation depth is 100%

the max. is approx..150%.

4-2-3-4.

4-2-3-4.

4-2-3-4.

4-2-3-4.

LINEARITY

LINEARITY

LINEARITY

LINEARITY

FREQUENCY

FREQUENCY

FREQUENCY

FREQUENCY

SWEEP

SWEEP

SWEEP

SWEEP

Press

MOD

key to select linearity sweep, and

Sweep

indicator light, the instrument is in linearity frequency sweep mode, press

f

0

key to select initial frequency, then, the

screen displays

0

******

”;

press

f

1

key to set terminate frequency, the screen displays

1

******

”;

press

∆f

key to set spacing frequency, the screen displays

2

******

”;

press

∆t

key to set sweep time interval

After finishing setting

∆t

”,

it begins to cycle sweep according to positive linearity. It is suitable for sweeping in a narrower frequency band.

For example, if you want to sweep between 1KHz to 1.1Khz by the frequency step value of 1Hz, set the time interval as 1s:

press

MOD

key to select linearity sweep mode

press

f

0

key and

1

,

KHz/mV

key to set initial frequency

press

f

1

key and

1

,

.

,

1

,

KHz/mV

key to set terminate frequency;

press

∆f

key and

1

,

Hz/%/S

key to set sweep spacing frequency

press

∆t

key and

1

,

%/S

key to set time interval

and then, sweep begins.

4-2-3-5.

4-2-3-5.

4-2-3-5.

4-2-3-5.

LOGARITHM

LOGARITHM

LOGARITHM

LOGARITHM

FREQUENCY

FREQUENCY

FREQUENCY

FREQUENCY

SWEEP

SWEEP

SWEEP

SWEEP

Press

MOD

key to select logarithm frequency sweep, and

Sweep

and

EXT/LOG

indicator light, the instrument is in logarithm frequency sweep mode

press

f

0

key to set

initiate frequency

press

f

1

key to set terminate frequency

press

∆t

key to set time interval. After finishing setting

∆t

, sweep begins. The setting way and displaying are the sa me as

that of linearity sweep, and it is unnecessary to set

∆f

. It is suitable for sweeping in a wider frequency band.

For example, if want to make logarithm sweep between 20Hz and 20KHz, time interval is 0.1s, take the following

operations:

Press

MOD

key to select logarithm sweep mode

Press

f

0

and press

2

,

0

,

Hz/%/S

key to set initial frequency

Press

f

1

key and press

2

,

0

,

KHz/mV

key to set terminal frequency

Press

∆t

key and press

0

,

.

,

1

,

%/S

key to set time interval, then, sweep begins.

EXPLANATION FOR FREQUENCY SWEEP

No matter which mode the frequency sweep is in, press any keys of

f

0

,

f

1

,

∆f

,

∆t

to stop sweeping, p ress

%/S

key to pause, then, the screen displays the frequency of t he

pausing point, and will output this value. When press

%/S

key again, it will go on sweeping. If need to restart, press

∆t

key and then

%/S

key. The sweep begins with finishing

input the unit of time interval

∆t

, so, time interval should be set lastly.

For

For

For

For

digital

digital

digital

digital

input

input

input

input

mode,

mode,

mode,

mode,

if

if

if

if

the

the

the

the

digit

digit

digit

digit

before

before

before

before

decimal

decimal

decimal

decimal

point

point

point

point

is

is

is

is

0

0

0

0

,,,

,

the

the

the

the

0

0

0

0

can

can

can

can

not

not

not

not

be

be

be

be

omitted.

omitted.

omitted.

omitted.

For any frequency output, TTL level square wave is in synchronism with sine output, and it is unnecessary to switch ranges; After power off or reset, the setting data on each

ranges become the default value (F=1KHz, V=1V, DC offset=±0.15V, AM modulation frequency=400Hz, AM modulation amplitude m=50%, sweep initial frequency f

0

=1KHz, sweep

terminal frequency f

1

=10KHz, sweep interval frequency ∆f=1KHz, sweep interval time ∆t=1s).

5.

5.

5.

5.

NOTES

NOTES

NOTES

NOTES

5-1.There should not exist DC and high voltage and strong radiance signal on the load of the generator to prevent from damaging the instrument.

5-2.When the instrument appears the problem, such as un-normal displaying, breaking down. etc. just press reset key or turn off the power, the problem will be solved.

5-3.If there is a lower DC in load circuit or DC component is unnecessary, you can joint-connect a capacitor with proper capacitance and withstand voltage to interclude DC.

5-4.It is better not to connect a oversize load on TTL level output, otherwise, it will lead to wave form distortion, long rising time.