Table 3-2. Voltage, Current, Ohms and Frequency Rang and Full Scale Value
FUNCITION RANG
FULL
SCALE INFORMATION
V
~
and V
300mV 330.00mV
*760V ac rms
3V 3.3000V
30V 33.000
V
300 V
330.00 V
1000V* 1100.0
V*
mA
~
and
mA
0.3mA** 0.33000
mA**
**AC current have not 0.3mA and
3mA range
3mA** 3.3000
mA**
30mA 33.000
mA
300 mA
330.00 mA
A
~
and
A
10A 11.000A
Resistance
300Ω
330.00
Ω
3K
Ω
3.3000
K
Ω
30 K
Ω
33.000
K
Ω
300 K
Ω
330.00
K
Ω
3M
Ω
3.3000
M
Ω
30 M
Ω
33.000
M
Ω
100 M
Ω
110
M
Ω
Frequency
300Hz 330.00
Hz
30MHz range have not full scale
3KHz 3.3000
KHz
30KHz 33.000
KHz
300KHz 330.00
KHz
3MHz 3.3000
MHz
30MHz
/
ANGING (AUTO, , )
Ranging operations are performed using the AUTO, and buttons (see Figure 3-4).
Autoranging
When you are in the autorange mode, the “AUTO” annunciator is lit.
In autorange, the meter automatically selects
the next higher range when a reading is
greater than full scale. If no higher range is
available, “OL” (overload) is displayed. The
meter automatically selects a lower range
when a reading is less than approximately
10% of the selected range.
Figure 3-4. Range Selection Butons
D C V
A C V
F R E Q
D C A
A C A
A U T O
M N
S H I F T
M X
H O L D
C O M P
O N
1
P O W E R
O F F
0
R A N G E
d B m
R E L
R A T E
r e a d i n g r a t e
A C A
F R E Q
D C A
A C V
D C V
V o l t s D C
A m p s D C
R e s i s t a n c e
V o l t s A C
A m p s A C
D i o d e
C o n t i n u i t y
F r e q u e n c y
d B m
d B m