16
• When the resistance above 1MΩ is measured, the
reading on the meter will not be stable until several
seconds have passed. This is normal for high-resistance
measurement.
• When the resistor is measured, do not input voltage
value. Never input the voltage above the overload
protection, otherwise, the meter may be damaged and
the operator may be hurt.
• After the measurement is over, immediately
disconnect the test probe from the tested circuit.
Resistance (Ω)Technical Indicators
Range
Accuracy
Resolution
400Ω
±(0.8%+5d)
0.1Ω
4kΩ
±(0.8%+4d)
1Ω
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
±(1.2%+10d)
10kΩ
Open voltage: 400mV
Overload protection: 250V DC or AC peak.