VICI CM-2016 Operation Download Page 10

10 

 

 

4.4.DCA MEASUREMENT 

(I) Turn the rotary function switch to A

 position. 

The meter is default at DC A measurement after 
power on. Press SELECT to switch 
between AC or DC measurement. 
(2) When the meter is used in a strong 
magnetic field, the readings could 
be unstahle or inaccurate. Press REL 
to clear to zero before measurement. 
(3)Press the clamp gunlock to open  
the clamp jaw to get testing wire  
inserted  in the middle and clipped 
 completely. 
To ensure measurement accuracy, please check to ensure that  
the testing wire is placed in the middle of clamp jaw (fig.5).  
Measure one wire in one time. If measure more than one wires 
in one tirne, the reading will be inaccurate. 

NOTE: 

(1) The meter uses Hall component to detect current. Hall 
component is a sensitive component. lt is magnetic sensitive. 
It is also sensitive to temperature and mechanical stress. 
Any impact on it would cause the changes on reading in a short time. 
This causes some remaining values  on the DC ampere reading 
when there is no current. Toget accurate reading, press REL key 
to clear reading before measurement, and the testing wire must 
be placed in the middle of the clamp jaw. 
(2) Press the clamp gunlock to open the clamp jaw to get testing 
wire inserted in the middle and clipped completely. Ensure that 
the testing wire is placed in the middle of clamp jaw. Otherwise 
there will be additional accuracy ±l.0% of reading. Read the 
measurement current value of the testing wire on the LCD. 
(3 )In the measurement of the DC current, if the reading is 
positive value, the current is top to down (broad is top, bottom 
cover is down). 
(4) The max.testing current is 600A. Exceeding the rated current 
for long time will damage the meter. 
Follow below instructions to get more accurate DC current 
measurement: 
1.) Switch off the testing wire. 
2.) Press the clamp gunlock to open the clamp jaw to get testing 
wire inserted in the middle and clipped completely. Ensure  that 
the testing wire is placed in the middle of clam pjaw. 
3.) When the meter reading is stabilized in the minimum value, 
press the REL key to clear it. 

Summary of Contents for CM-2016

Page 1: ...1 DIGITAL CLAMP METER OPERATION...

Page 2: ...jury caused by wrong operation please read this manual carefully before operation 1 SUMMARY This is a 3 5 6 digit digital AC DC clamp meter with good stability and reliability It has a LCD with 14mm d...

Page 3: ...g operation 2 2 2 Disconnect all signals and plug offtest leads before opening the case cover To avoid electric shock or damage to the meter water should not be leaked inside 2 2 3 DO NOT use the mete...

Page 4: ...humidity 70 3 1 15 Dimension 123Wx270Dx35H mm 3 1 16 Weight approx 280g including batteries 3 1 17 Accessories operation manual x 1pc test leads x 1 pair temperature probe TP01 x 1 pair 1 5V AAA batte...

Page 5: ...curacy Resolution 600 A 2 0 30 0 l A Frequency response Sine wave and triangle wave 40Hz 1kHz other waveforms 40Hz 200Hz NOTE The testing object shall be placed in the middle of the clamp jaw 3 2 7 Re...

Page 6: ...40 C 1000 C 1 0 5 400 C 1 C 1 5 15 400 C 0 F 1832 F 0 75 5 750 F 1 F 1 5 15 750 F Sensor K type banana plug Nickel Chromium Nickel Silicon WARNING DO NOT input voltage in this range 3 2 11 Diode and...

Page 7: ...ement Press Hz DUTY key to switch measurement between frequency and duty ratio in low voltage amplitude C F Temperature measurement Press SELECT key to switch between C and F 7 FUNCTlON KEYS SELECT ke...

Page 8: ...LCD panel display the measurement values and units No Function Description 1 MAX MIN Invalid 2 APO A symbol of auto power off 3 Clear to zero or in REL measurement mode 4 HOLD Data hold funcions is a...

Page 9: ...cuit to avoid electric shock DO NOT touch the circuit by hands or bodies 3 Disconnect test leads from the testing circuit when all tests are finished 4 3 ACV MEASUREMENT I Turn the rotary function swi...

Page 10: ...some remaining values on the DC ampere reading when there is no current Toget accurate reading press REL key to clear reading before measurement and the testing wire must be placed in the middle of th...

Page 11: ...ax testing current is 600A Exceeding the rated current for long time will damage the meter 4 6 RESISTANCE MEASUREMENT I Turn the rotaryfunction switch to position Press SELECT to choose measurement 2...

Page 12: ...play OL If there is breakdown of creepage on the diode the LCD display a certain value 4 A complete diode testing includes forward and backward measurements If the result doesn t meet the descriptions...

Page 13: ...UTY key to switch between frequency and duty ratio measurements NOTE I When the input exceeds 10Vrms change to ACV range and then press Hz DUTY key to switch between frequency and duty ratio measureme...

Page 14: ...input in the meter s terminal the voltage indicator may light on because of voltage sensor installed in the meter 4 Interference sources in the external environment such as flashing light motor would...

Page 15: ...ork properly please check the mete ras following steps If the problems remain please contact repairing center or local dealers Fault Solution No reading on LCD Turn on the power Release the HOLD key R...

Reviews: