5.8. FUSES AND VOLTAGE SETTINGS
Fuse values depend on voltage and device settings. Use only 5x20mm fuse slow blow.
INSTRUMENT SETTINGS
INPUT VOLTAGE
SETTING (V AC)
FUSES (A T)
PARTS
FUSE MARK
Hydrogen Part
F3
110
6.3
220
4
Zero Air Part 1500
F2
110
2
220
2
Zero Air Part >1500
F2
110
6.3
220
6.3
Check the setting of the voltage selector on the rear of the unit. The set voltage is indicated in the white
square.
• Use only 240 V AC or 120 V AC settings and change fuses if needed.
• Turn off the instrument before changing the voltage or fuse.
To change the voltage or fuse, proceed as follows:
Use a small flat screwdriver to push the snap clip of voltage selector insert to
remove it.
B. To change the fuse value
, just replace it. Use a screwdriver to unscrew the
cap from fuse holder F3. Replace the fuse using a 5x20mm fuse, and screw the
cap into its original position.
A. To change the voltage
, spin the white
holder. Re-insert the voltage selector and
check to assure the white window shows the
desired voltage.