The VICI DBS® Mistral Evolution Hybrid is a self contained generator that produces up to 12 L/min of high 
purity nitrogen and 22 L/min of air. The generator is designed to meet the specific requirements of the 
Sciex LC/MS. Nitrogen is produced by utilizing a combination of compressor and Carbon Molecular Sieve 
(CMS) technology. High and low pressure compressors are carefully matched to the CMS demand to ensure 
quite and reliable operation. This unique combination of dual compressor technology has several unique 
advantages over all other nitrogen generators commercially available. 

2.1 Symbol Definitions

SYMBOL

DESCRIPTION

General warning. Caution is necessary when operating the device or control 

close to where the symbol is placed, or to indicate that the current situation 

needs operator awareness or operator action in order to avoid undesirable 

consequences.

The manual must be consulted in all cases where this symbol is marked on 

the device.

Risk of asphyxia. Caution is necessary when operating the device in a not 

sufficiently ventilated area.

Oxidant material. Caution is necessary due to the presence of oxidant material.

WEEE symbol. When discard this product, it must be sent to separate collection 

facilities for recovery and recycling.

Fuse. Notify replaceable fuse place.

OR

OR

OR

Summary of Contents for MISTRAL EVOLUTION HYBRID

Page 1: ...ort vicidbs com USA Canada Central South America techsupportusa vicidbs com Italy service vicidbs com VICI AG INTERNATIONAL Parkstrasse 2 CH 6214 Schenkon Switzerland tel 41 41 925 6200 fax 41 41 925...

Page 2: ...iew 6 5 3 Connections 7 5 4 Power Supply Settings 7 5 5 Transport and Storage 7 6 SCHEMES 8 6 1 Fluidics 8 6 2 Electrical 9 7 STARTING THE GENERATOR 10 8 SOFTWARE 11 8 1 Overview of the Menus 11 8 2 P...

Page 3: ...N HISTORY VERSION DATE COMMENT CHANGE 1 00 0100 01 10 2017 Layout and initial release VICI DBS 1 01 0000 12 20 2018 Update 1 02 0000 05 18 2020 New layout images and graphics VICI Gig Harbor Group 1 0...

Page 4: ...veral unique advantages over all other nitrogen generators commercially available 2 1 Symbol Definitions SYMBOL DESCRIPTION General warning Caution is necessary when operating the device or control cl...

Page 5: ...y time If the Mistral Evolution is used in a way not specified by VICI DBS the protection provided by the equipment could be compromised 3 2 Instrument Safety Instructions The Mistral Evolution Hybrid...

Page 6: ...barg psig 7 100 Technology Carbon molecular sieve Warm up time minutes 20 Electrical supply 110 120V 60Hz 220 240V 50Hz Power consumption watts 1800 Noise level 55 dBA 1 meter Dimensions mm in 482W x...

Page 7: ...w Command Buttons x3 LED Indicator Lights Backlit Display 5 2 Back View Relay Alarm Outlet Compressor Box Cooling System Dust Silencer Filter Outlet Water Purge Outlet N2 with Valve Outlet Pressure Se...

Page 8: ...talled to transform internally the requested voltage to 110 115 V 50 60Hz The list of voltage and frequencies settings possible are bellow In any case the voltage of the power supplied must stay at 10...

Page 9: ...back to the factory for recalibration or repair the Mistral Evolution Hybrid should be packed in its original packing with the inlets and outlets protected against any accidental penetration of dust o...

Page 10: ...S DOCUMENT AND THE INFORMATION WHICH IT CONTAINS SHALL NOT BE USED EXPLOITED OR SOLD AND SHALL NOT BE REVEALED OR DISCLOSED TO OTHERS WITHOUT THE EXPRESS WRITTEN PERMISSION OF VICI THIS DOCUMENT SHALL...

Page 11: ...Mistral Evolution to an external water tank 4 Turn the generator on switch in back 5 Press the MENU button and scroll until you get the PARAMETER menu Press the SELECT button 6 Go into each sub menu...

Page 12: ...perature C F C Flow Unit Sets desired unit of measure for flow rate L min ft3 min L min Auto Start Choose to automatically start production upon powering on or not Yes No No Display Contrast Sets desi...

Page 13: ...situation it does not make sense to have the Mistral Evolution run continuously and throw away the produced Nitrogen The Mistral Evolution is able to handle an external tank management The Mistral Ev...

Page 14: ...ight size of the tank will be around 225 L 9 2 Where to Get the Tank As the size of the tank will be large for application and in addition legislation is different from one country to another DBS does...

Page 15: ...0 h service kit 10 ALARMS PRE ALARMS 10 1 Pre Alarms Precursors of the alarms Do not stop the device ITEM DESCRIPTION Temperature The temperature inside the cabinet is over the maximum limit Check if...

Page 16: ...ure is too high and do not cool the compressors The compressors are overheating Check and eventually replace the compressors Int tank pressure min The pressure in the internal reservoir is too low The...

Page 17: ...intenance Every 8 000 10 000h the compressors must be replaced DB EVO 002 OF 302 Air compressor hybrid and gas station 13 SPARE PARTS VICI DBS P N DESCRIPTION DB EVO 001 Complete Heat Exchanger DB EVO...

Page 18: ...66 Adapter for AF20 to FID S FID T Rack N2 Tower DB FIL010 Maintenance kit 4000h for MISTRAL EVO includes 1 white sil 3 blue sil 1 cart DB 10617 Outlet N2 on off Valve G1 4 for Mistral DB 10620 Silenc...