Disinfecting your Vicair Active O2
Warning!

 In case of soiling with body fluids, and 

when transferring the product to another user, 
the product must be disinfected to avoid cross 
contamination. Clean the Active O2 cover, foam 
inlay and SmartCell liners as instructed above. 
Preferably use an anti-bacterial detergent. After 
cleaning spray your cover, foam inlay liner and 
SmartCell liners with a disinfectant. Bend the 
cover compartments outwards to make sure the 
spray gets into the gaps.

(Re)assembling your Vicair Active O2

The Vicair Active O2 is fully assembled on 
delivery. When it needs to be (re)assembled, for 
instance after cleaning, follow the assembly 
instructions underneath. The instructions 
correspond to image i (I, II & III) on the Quick 
Installation Guide.

Placing the foam inlay in the Vicair Active O2 
cover (image i - I)
Caution! 

Do not use the foam inlay without the 

foam inlay liner.
•  Completely open the zipper at the front of 

the Active O2 cover.

•  Place the foam inlay in the Active O2 cover, 

the sloping side pointing to the back of the 
cushion and the contoured side on top.

•  Close the zipper. Make sure you do not 

damage the foam inlay liner.

Adjusting the number of SmartCells in the 
SmartCell liners (image i - II)
Warning!

 Vicair B.V. recommends to adjust the 

filling grade under the direct supervision of 
your therapist or Assistive Technology 
Professional.

The number of SmartCells in each SmartCell 
liner depends on the size of the cushion. You 
can find the default filling grade on the label 
inside the cushion cover. When it’s necessary to 
adjust the number of SmartCells, follow these 
steps:
•  Open the SmartCell liner’s zipper and add or 

remove the desired number of SmartCells.

•  Close the zipper. Make sure that you do not 

damage the SmartCells.

Placing the SmartCell liners in the Vicair Active 
O2 cover (image i - III)
Warning!

 Make sure the SmartCell liners are 

placed in the correct compartment (see 
SmartCell liner label).
•  Completely open the zipper at the back of 

the cushion cover.

•  Place the SmartCell liners in the cover. Make 

sure you place the correct SmartCell liner in 
each compartment. The SmartCell liner’s 
label states where it should be placed.

•  Close the zipper. Make sure that you do not 

damage the SmartCell liners.

REPORTING OF SERIOUS INCIDENTS

Serious incidents from use of Vicair products 
should be reported to the manufacturer and the 
relevant competent authority.

WARRANTY

Please see www.vicair.com for complete 
information on warranty processes, terms and 
conditions.
Vicair B.V. offers a limited warranty against 
defects in workmanship and materials, 
provided the product has been used normally. 
Any damage arising from abnormal use, or 
caused by improper handling, cleaning, or 
negligence, is excluded from this warranty.

Summary of Contents for Active 02

Page 1: ...USER MANUAL ...

Page 2: ...EIDING 3 GEBRAUCHSANWEISUNG 4 MANUEL DE L UTILISATEUR 5 BRUGERVEJLEDNING 6 BRUKSANVISNING 7 BRUKERMANUAL 8 KÄYTTÖOHJE 9 MANUAL DEL USUARIO 0 MANUAL DE UTILIZADOR q MANUALE UTENTE w ΕΓΧΕΙΡΙΔΙΟ ΧΡΗΣΗΣ e KORISNIČKI PRIRUČNIK r UPORABNIŠKI PRIROČNIK t INSTRUKCJA OBSŁUGI y 用户手册 u ユーザーマニュアル ...

Page 3: ...ive O2 allows for easy transfers and provides stable body support comfort and helps to prevent pressure injuries It is also a high performance addition to a pressure injury treatment plan stage I IV CONTRAINDICATIONS When used as intended there are no known contraindications MAIN COMPONENTS A B C D E A Foam inlay with fluid resistant foam inlay liner placed in front compartment B Vicair Active O2 ...

Page 4: ...liner label Warning Vicair B V recommends to adjust the filling grade under the direct supervision of your therapist or Assistive Technology Professional Warning In case of soiling with body fluids and when transferring the product to another user the product must be disinfected to avoid cross contamination Warning Every three months check the product for wear and deflated cells In case of serious...

Page 5: ...g out Image 6 Quick Installation Guide Warning Your tailbone and pelvic bones should not touch the bottom of the cushion Checking for bottoming out should be done during first use and after adjustments to the cushion filling Warning Make sure that your tailbone and pelvic bones are resting on the rear compartments with SmartCell liners not on the foam inlay Warning Vicair B V recommends to adjust ...

Page 6: ...e wheelchair and into a car manufacturer installed vehicle seat without the cushion and use the vehicle s crash tested occupant restraint system As an alternative for your Vicair wheelchair cushion the Vicair Allrounder O2 which can be fastened to the user s body can be used if you require additional protection in a manufacturer installed vehicle seat during transportation When the wheelchair is u...

Page 7: ...nst the glass of the washer door Caution Excessive washing may influence the aging of this product Cleaning the Vicair Active O2 After you have emptied the storage pouch and removed the foam inlay your Vicair Active O2 cushion can be machine washed at 60 C 140 F using an anti crease cycle and mild detergent You do not need to remove the SmartCell liners before putting the cushion into the washing ...

Page 8: ...ing grade under the direct supervision of your therapist or Assistive Technology Professional The number of SmartCells in each SmartCell liner depends on the size of the cushion You can find the default filling grade on the label inside the cushion cover When it s necessary to adjust the number of SmartCells follow these steps Open the SmartCell liner s zipper and add or remove the desired number ...

Page 9: ... period Vicair B V will at its own choice repair or replace the product free of charge Remedies for breach of express warranties herein are limited to repair or replacement of the product In no event shall damages for breach of any warranty include any consequential damages or exceed the cost of non conforming goods sold Any questions Contact Vicair B V at info vicair com or contact your country s...

Page 10: ...rgt voor een gemakkelijke verplaatsing en biedt een stabiele lichaamsondersteuning comfort en helpt decubitus te voorkomen Het is tevens een hoogwaardige toevoeging aan een decubitusbehandelplan graad I IV CONTRA INDICATIES Bij gebruik volgens de instructies zijn er geen contra indicaties bekend BELANGRIJKSTE ONDERDELEN A B C D E A Schuimblok met vloeistofbestendige schuimblokvoering dat zich in h...

Page 11: ...rs in het juiste compartiment worden geplaatst zie SmartCell liner etiket Waarschuwing Vicair B V raadt aan om de vulgraad aan te passen onder direct toezicht van uw therapeut of hulpmiddelenadviseur Waarschuwing In geval van vervuiling met lichaamsvloeistoffen en bij het gebruik van het product door een andere gebruiker moet het product gedesinfecteerd worden om kruisbesmetting te voorkomen Waars...

Page 12: ...t stevig aan Zorg ervoor dat het schuimblok zich aan de voorzijde van de rolstoelzitting bevindt Ga enkele minuten op het kussen zitten zodat het zich aan de vorm van uw lichaam kan aanpassen Zorg ervoor dat uw staartbeen en bekken op de achterste compartimenten met SmartCell liners rusten en niet op het schuimblok Pas indien nodig de voetsteunen aan Controleren op doorzitten Afbeelding 6 Quick In...

Page 13: ... is namelijk gevuld met lucht waardoor het kussen zich anders kan gedragen op grote hoogte of in een vliegtuig Onder dergelijke omstandigheden kan de stevigheid van het kussen worden aangepast door enkele cellen uit elk compartiment te verwijderen Luchtvochtigheid Minimum 0 maximum 100 Dagelijks opschudden Vicair B V beveelt aan om uw Vicair Active O2 kussen dagelijks op te schudden voor gebruik G...

Page 14: ...s zeer groot dat het product en of de hoes van het product sneller slijten Voorzichtig Gebruik geen agressieve of schurende schoonmaakmiddelen of scherpe voorwerpen om het kussen te reinigen Voorzichtig Verwijder het schuimblok voordat u het kussen in de machine wast Opgelet Het schuimblok niet zonder de schuimblok liner gebruiken Voorzichtig Plaats de bovenkant van het Vicair O2 kussen niet tegen...

Page 15: ...k in de Vicair Active O2 hoes plaatsen Zie afbeelding i I Opgelet Het schuimblok niet zonder de schuimblok liner gebruiken Open de rits aan de voorzijde van de Active O2 hoes volledig Plaats het schuimblok in de Active O2 hoes met de schuine zijde gericht naar de achterkant van het kussen en de gegolfde kant aan de bovenzijde Sluit de rits Zorg ervoor dat u de schuimblok liner niet beschadigt Het ...

Page 16: ...hanteren reinigen of nalatigheid valt niet onder deze garantie Garantieperiode Vanaf originele aanschafdatum 24 maanden De SmartCells in dit Vicair O2 kussen hebben een levenslange garantieperiode Indien er binnen bovengenoemde garantieperiode productie of materiaalfouten optreden zal Vicair B V naar eigen inzicht en goeddunken het product kosteloos repareren dan wel vervangen De garantie beperkt ...

Page 17: ...ahrern mit hohem Risiko einer Bildung von Druckverletzungen ein besonders hohes Mass an Hautschutz bietet Es ist ausserdem eine hervorragende Ergänzung zu einem Druckverletzungs Behandlungsplan Stufe I IV INDIKATIONEN Das Vicair Active O2 erlaubt einfache Transfers bietet Sitzkomfort und eine stabile Stütze für den Körper und hilft Druckverletzungen zu vermeiden Es ist ausserdem eine hervorragende...

Page 18: ...nte försvinner inom 15 minuter Varning Se till att SmartCell liners placeras i korrekt fack se SmartCell liner etiketten Varning Vicair B V rekommenderar att justering av fyllnadsgraden sker under direkt överinseende av en terapeut eller en hjälpmedelskonsulent Varning Om produkten smutsas ner med kroppsvätskor och när den överförs till en annan användare måste den desinficeras för att undvika kor...

Reviews: