Viavi SmartPocket OLP-37XV2 Operating Manual Download Page 15

OLP

-

37XV2

15

P

OWER

 S

UPPLY

3  G

ETTING

 S

TARTED

Battery operation

Replacing the batteries

The battery compartment is on the back of the device.

1.

Pull down the lid to open the battery compartment.

NOTICE: Take care to insert the batteries correctly. 

The correct polarity is indicated by a diagram inside the battery 

compartment.

2.

Insert new batteries or replace dead ones.

3.

Close the battery compartment.

4.

Press [

E

] to switch on.

NOTE:

The batteries cannot be recharged with the OLP

-

37XV2.

General tips on using batteries

• Never use batteries based on lithium.

• Always handle batteries with care.

• Do not drop or damage the batteries or expose them to 

excessively high temperatures.

• Do not store rechargeable batteries for more than one or two 

days at very high temperatures (e.g. in a vehicle), either 

separately or fitted in the device.

• Do not leave discharged batteries in the device for a long time 

if it is not being used.

• Do not store rechargeable batteries for more than 6 months 

without recharging them at intervals.

• Avoid deep discharging of the batteries as this can cause the 

cell polarity to reverse and make the battery useless.

 WARNING

Dangers in handling batteries

Handling batteries may be dangerous. Please note the 

following safety instructions.

J

Please note the battery operation safety information in the 

section 

“Battery operation” on page 10

.

Replacing the batteries

J

Do not replace individual batteries. Always change both 

batteries at the same time.

J

Always use batteries of the same type; i.e. do not mix 

rechargeable and non-rechargeable batteries.

Summary of Contents for SmartPocket OLP-37XV2

Page 1: ...OLP 37XV2 SmartPocket Optical Power Meters Operating manual BN 2335 98 11 2021 01 English ...

Page 2: ...ute ETSI These recommendations and standards are subject to the proprietary rights of these organizations It is not permitted to copy ITU T recommendations or ETSI standards fully or in part and or to pass them on to third parties without prior written permission from ITU T and or ETSI Copyright Copyright 2021 Viavi Solutions Inc All rights reserved Viavi and the Viavi logo are trademarks of Viavi...

Page 3: ...iew 13 Power Supply 14 OPERATION 17 Switching the device on off 17 Display elements 18 Settings menu 19 Selecting a operation mode 20 Selecting a display mode 21 Absolute power level mode 21 Loss mode 22 Pass Fail mode 23 MEMORY MANAGEMENT 24 General information 24 Storing measurements 24 Recalling measurements 24 Deleting measurements 25 DATA EXPORT AND FIRMWARE UPDATE 26 USB and SmartReporter 26...

Page 4: ...4 OLP 37XV2 CONTENTS SPECIFICATIONS 29 OLP 37XV2 29 General specifications 29 ORDERING INFORMATION 31 Devices 31 Accessories 31 ...

Page 5: ... PC based reporting all test results can be summarized in a professional industry proven report Main features The OLP 37XV2 offers many helpful features that are optimized to workflows of typical telecom and cable TV operators and thus ensure that test times are kept as short as possible Accurate and repeatable wavelength selective power measurements Unambiguous pass fail result presentation and u...

Page 6: ... provided by your device are missing please visit the Viavi web site to check if additional information is available To download the latest operating instructions 1 Visit the Viavi web site at www viavisolutions com 2 Search for SmartPocket V2 3 Open the download area and download the operating instructions if available ...

Page 7: ...ment CAUTION Follow the instructions carefully to avoid a low or medium risk of injury to persons WARNING Follow the instructions carefully to avoid potential death or severe injury to persons DANGER Follow the instructions carefully to avoid death or severe injury to persons High Voltage Follow the instructions carefully to avoid damage to the instrument or severe injury to persons This safety in...

Page 8: ...J 1 2 Instruction Follow the instructions given the numbers indicate the order in which the instructions should be followed e g J Select mode Italics Result Indicates the result of following an instruction e g The page opens Bold type face Pages controls and display elements Screen pages controls and display elements are indicated in bold type Text in blue Cross references Cross references are ind...

Page 9: ... ambient conditions J Observe the specified measurement range J Always make sure that the instrument is in proper working order before switching it on Warning symbols indicating a potential hazard J A warning symbol on the device indicates a potential hazard In all cases where the a warning symbol is shown on the display or labeled on the device the operating manual must be consulted to learn more...

Page 10: ... Switch off the laser source before disconnecting the optical fibers J Never look directly into the output of a laser source or into an optical fiber connected to it J Always cover unused ports J Heed the normal precautions for working with laser radiation and consider local regulations WARNING Explosion danger Short circuiting the batteries can result in overheating explosion or ignition of the b...

Page 11: ...es not have a recharge function for rechargeable batteries so there is no danger when using dry batteries J Read the manual of the external charging device NOTICE Insufficient ventilation Insufficient ventilation can damage the device or adversely affect its function and safety J Ensure adequate ventilation when operating the instrument ...

Page 12: ...damaged during shipping This is particularly likely if the packaging is visibly damaged If there is damage do not attempt to operate the device Doing so can cause further damage In case of damage please contact your local Viavi Sales Company Addresses can be found at www viavisolutions com Recovery following storage shipping Condensation can occur if a device that is stored or shipped at a low tem...

Page 13: ...2 Fixed SC adapter 3 Display 4 Key pad Representation in the user manual Context sensitive keys here left key is selected MODE Mode Settings key G Save Results key E Power key 5 USB interface For power supply and measurement data downloads and updates 6 Battery compartment on rear of the device ...

Page 14: ... press Holding the key for at least 2 sec opens the second level You then have access to additional functions or a menu to change settings Power Supply NOTE The devices are not designed for batteries based on lithium The following power sources can be used to operate the OLP 37XV2 Two 1 5 V dry batteries Mignon AA size alkaline type recommended Two 1 2 V NiMH rechargeable batteries Mignon AA size ...

Page 15: ...t store rechargeable batteries for more than one or two days at very high temperatures e g in a vehicle either separately or fitted in the device Do not leave discharged batteries in the device for a long time if it is not being used Do not store rechargeable batteries for more than 6 months without recharging them at intervals Avoid deep discharging of the batteries as this can cause the cell pol...

Page 16: ...trash You will often be able to return used batteries to the place where you purchase new ones Any dry or rechargeable batteries that you purchased from Viavi can be returned to one of our Service Centers for disposal Operation from AC power To fit one of the mains plug adapters J See Fig 1 and follow the instructions which are shown on the packaging of the mains plug adapter Fig 1 Fitting the mai...

Page 17: ...To change the power mode I The device is switched on 1 Long press MODE to open the settings menu 2 Use to select ECON 3 Press to select power mode ON ECON OFF PERM 4 Press MODE to close the menu Mode Icon Description Permanent ON PERM The device is switched on permanently Automatic OFF ECON The device switches off 20 minutes after the last operation This function is only available when the device ...

Page 18: ...n be changed via the settings menu Battery status in PERM power mode Device remains switched on Battery status in ECON power mode Device switches off 20 minutes after the last operation The device is powered via USB 2 Center of display db dBm Shows the measurement results in dBm dB or W PON nm Display of selected mode and wavelength 3 Context sensitive key functions 1490nm Dual Selection of mode 1...

Page 19: ...change setting Press once to change one step at a time Hold down the key to increase the step change rate Minute 1 59 Year 2020 2030 Month 01 12 Day 01 31 About Show device data including last calibration date J Press to confirm setting J Press to return to the settings menu Factory Reset Reset Auto Lambda ON OFF Press right context key to toggle ON OFF Bluetooth LE ON OFF ECON ON OFF ON ECON OFF ...

Page 20: ... operation modes with following displays G PON Gigabit Passive Optical Network measurement at 1490 nm XGS PON 10 Gigabit Capable Symmetric Passive Optical Network measurement at 1577 nm Dual Display of G PON and XGS PON simultaneously To select a operation mode J Press to select a mode Fig 2 Operation modes ...

Page 21: ...ative to a reference value PF Abs Pass Fail indication based on an absolute power threshold PF Loss Pass Fail indication based on a relative power threshold To select a display mode J Press to toggle between the displays modes Absolute power level mode The power level is displayed in dBm or Watts nW µW mW J Long press to toggle dBm Watts NOTE The context sensitive field shows the selected unit ...

Page 22: ...e reference value Abs Ref J Press to display relative power level Setting the reference level I Absolute power level mode or Loss mode is selected J Press Abs Ref The current power level is set as the new reference level NOTE The reference level can be stored for both wavelength separately When Dual is selected the power levels of both wavelengths are set as reference level simultaneously ...

Page 23: ...ode J Press until PF Abs or PF Loss is displayed NOTE When PF Abs or PF Loss is selected the two modes can also be toggled by pressing Setting the fail threshold Setting the fail threshold is identical for both Pass Fail modes I PF Abs or PF Loss mode is selected Press to toggle between both modes 1 Press to select a wavelength see also page 20 2 Long press The Set Fail Threshold screen opens 3 Us...

Page 24: ...asurements J Press G to save the current result The result is saved when Saving appears on the display and below the name of the currently saved data Theresults arealways stored with the current date timestamp e g 2021 01 13T16 05 52 corresponding to January 13 th 2021 at 16h 05min 52s The new results are always appended successively at the last memory location even if you clear a previously assig...

Page 25: ...eting measurements I The device shows the list of saved measurements 1 If you want to delete a single entry use to select it 2 Long press to open the sub menu 3 Press to delete the selected entry or press to delete all results A dialog will aks you to confirm the deletion 4 Press to accept or press X to cancel 5 Press to close the sub menu and to show the list of saved measurements or press MODE t...

Page 26: ...pdate the firmware The SmartReporter reporting tool is offered by Viavi for free and can be download from https updatemyunit net Application Software J For more information about data export and firmware update via USB and SmartReporter please refer to the SmartReporter user manual Bluetooth and MobileTech app StrataSync cloud You have also the choice to upload the saved data and to update the fir...

Page 27: ...tions with the dust cap whenever they are not in use This prevents them from getting dirty WARNING Invisible laser radiation Maintenance or cleaning of the instrument while it is connected or operating may damage the instrument or injure you J Make sure that the instrument is switched off and disconnected from all power sources and optical radiation sources before maintenance or cleaning J Do not ...

Page 28: ...ts dirty through use you can clean it using a soft cloth moistened with a mild solution of detergent NOTICE Water and cleaning fluids The instrument may be damaged or destroyed if water or cleaning fluids penetrate it J Make sure that water or cleaning fluids do not get inside the instrument ...

Page 29: ...s band 1540 to 1650 nm Power uncertainty 0 5 dBm1 1 At 2 dBm at 23 C 3 C at nominal wavelength Calibrated wavelengths 1490 nm 1577 nm Optical adapter system Fixed SC Memory capacity 1000 measurement results Data readout via USB C interface Recommended calibration interval 3 years Dry batteries 2 x AA 1 5 V never use batteries based on lithium Rechargeable batteries NiMH 2 x AA 1 2 V Power consumpt...

Page 30: ...tromagnetic compatibility EMC EN 61326 1 2013 Device safety EN 61010 1 2010 Operating temperature range 10 to 55 C 14 to 131 F Storage and transport 40 to 70 C 40 to 158 F Altitude 2000 m max 6500 ft max Pollution degree 2 Relative humidity up to 31 C 15 to 85 Absolute humidity 31 C 1 to 29 g m3 Dimensions H x W x D 30 x 80 x 150 mm 1 18 x 3 15 x 5 90 in Weight incl batteries 200 g 0 44 lb ...

Page 31: ...terials power supplies Optical Power Meter G PON XGS PON OLP 35XV2 OLP 37XV2 BN 2333 90 03 OCK 10 Optical cleaning kit BN 2229 90 21 Cleaning tape for optical connectors BN 2229 90 07 Spare optical cleaning tape BN 2229 90 08 NiMH rechargeable batteries Mignon AA 1 2 V 2 batteries required BN 2237 90 02 AC adapter BN 2302 90 01 ...

Page 32: ...design and manufacture Environmental restrictions and requirements are taken into account during planning and manufacture of Viavi products This attention ranges from the raw materials and finished components selected for use and the manufacturing processes employed through to the use of energy in the factory and right on up to the final stages in the life of a product including dismantling Hazard...

Page 33: ...ion EC No 1907 2006 requires article suppliers to provide information if a listed Substances of Very High Concern SVHC is present in an article above a certain threshold For information on the presence of REACH SVHCs in Viavi products see the Hazardous Substance Control section of Viavi s Standards and Policies web page EU CE Marking Directives LV certified by TÜV SÜD EMC RoHS RE This product conf...

Page 34: ...OLP 37XV2 34 PRODUCT REGULATORY COMPLIANCE ...

Page 35: ...OLP 37XV2 35 PRODUCT REGULATORY COMPLIANCE ...

Page 36: ...er Meters OLP 37XV2 viavisolutions com Viavi product specifications and descriptions in this document are subject to change without notice 2021 North America 1 844 468 4284 Latin America 1 954 688 5660 China 86 21 6859 5260 Germany 49 7121 86 0 ...

Reviews: