Chapter 15
Using Cable and Antenna Analyzer
530
JD700B Series User’s Guide
Figure 321 Connection for 2-port vector measurement
CAUTION
The maximum input power for the
Cable and Antenna Analyzer RF In
port is +25 dBm.
However, a proper level of input to the
Cable and Antenna Analyzer
RF In
port is
0 dBm. When the input power is expected to be greater than 0 dBm, you must adjust the
output power for the RF Out.
Connecting a cable for 2-port scalar measurements
Procedure
1
Connect the one end of the Thru test cable to the Cable and Antenna Analyzer Reflection/RF Out
port of the instrument and the other end of the cable to DUT.
2
Connect the DUT to the
Spectrum Analyzer RF In
port of the instrument as illustrated in
. Measurement result is displayed on the screen.
Figure 322 Connection for 2-port scalar measurement
CAUTION
The maximum input power for the
Spectrum Analyzer RF In
port is +25 dBm for JD780B
series and +20 dBm for JD740B series.