SHUNT CALIBRATION

 

The Model 118 is equipped with an external shunt calibration.  This feature allows field calibration to be performed by 
electrically simulating pressure.  When the shunt calibration is activated, the transducer bridge is unbalanced by an amount 
equivalent to the application of approximately 80% of full scale pressure.  The exact pressure and output values are listed 
on the Performance Certificate.  

External shunt calibration is achieved by placing the specified resistor across Pins 2 and 4 of the electrical connector.  The 
value of the resistor is listed on the Performance Certificate.  Internal shunt calibration is optional on the Model 118.  Units 
with internal shunt calibration have a 6 pin connector.  The shunt calibration resistor is built into the transducer and the 
circuit is activated by connecting (shorting) Pins 5 and 6 of the connector.  

After calibration, the shunt calibration circuit should be deactivated for normal operation. 

 

FIELD CALIBRATION

 

The Model 218 Series Transducers are provided with a unique feature that permits field calibration to be performed without 
the need of a pressure source, by electronically simulating pressure.  The nylon access screws should be removed and the 
pressure port shall be vented to zero pressure.  For transducers measuring absolute pressure, a full vacuum must be 
applied to the pressure port.  Connecting (shorting) the calibration terminals (Pins 5 and 6), produces an output equivalent to 
approximately 80% of full-scale pressure.  Refer to the performance certificate, which was supplied with the transducer, for 
the exact calibration output.  If the certificate is unavailable, the calibration output can be calculated by using the following 
formula: Volts DC = (Cal pressure from transducer label / pressure range) x 5.  Thus, following adjustment of zero output for 
zero pressure, the calibration circuit may be activated and the gain or span adjustment screw can be turned until the correct 
calibration output appears at the output terminals.  The calibration terminals should then be disconnected and the zero 
adjustment reconfirmed.  

Note - the calibration circuit shall be deactivated during normal operation.  The nylon cover screws, 

removed to access zero and span adjustments, shall be reinstalled after calibrating the transducer. 

 

 

 