healthcare
facilities.
SpO2
measurement
technology is based on developed photoelectron
method, the circuit design and calculation software
was developed by Shenzhen Viatom Technology
Co., Ltd. The SpO2 sensor receives the optical
signal from the red light and infra-red light through
the finger. Insert the finger into the oximeter, there
are two emitting tube (red light diodes and infrared
diodes) located on the inner upside of the sensor
and they can emit red light and infrared; There is
the receiving end located on the inner downside of
the sensor, and it can transmit the red light and
infrared into the pulse signal through finger. The
MCU receives the pulse signal, gets the frequency
signal by counting, processes its digital signal, and
finally gets the measured SpO2 value. The PR is
averagely calculated by above peak intervals of PR
waveform.
3.5
Stop working & Upload data
Take off the device, the countdown
will begin.
(If the working time is less than 2 minute, there will
be no countdown)
During the countdown, if you wear the device again,
the record will be resumed.
After the countdown
, the
data will be ready for
uploading
.
Notice:
The built-in memory can store 4 records.
The oldest will be overwritten by the 5th. Please
upload data to your phone in time.
3.6
Screen Wake up
The screen will go off automatically for saving
power in Standard Mode; you can touch the key on
top to wake up the screen.
3.7
How to Check Battery
Touch the key on top, you can switch display
between readings and battery.
3.8
Unavailable Symbol
When this symbol displays on device screen, it
indicates the readings is unavailable right now.
It may caused by:
Excessive movement;
Poor signal, finger is too cold;
Usually, the readings will recover in a few seconds
when at rest.
3.9
Bluetooth Connection
The device Bluetooth will be enabled automatically
after it’s turned on.
To establish a Bluetooth connection,
1)
keep the device Bluetooth enabled.
2)
Make sure the phone Bluetooth is enabled
.
Notice:
DO NOT PAIR
in the settings of your phone.
3.10 Add a New Device
For the initial use, you need to add a new device.
1)
Turn on device, run phone, select <O2Ring>;
2) Touch the key on top of device.
3.11 Smart Vibration by SpO
2
The vibrator in the sensor will be activated when
the SpO
2
fall below the pre-set value (Threshold).
The vibration will stop when the SpO
2
recover.
3.12 Smart Vibration by Pulse Rate
When your pulse rate is above the Upper
Threshold or below the Lower Threshold, vibration
will be activated.
4
Maintenance
4.1
Time & Date
After connection with phone, device time will
upload from your phone time automatically.
4.2
Cleaning
The device can be repeatedly used. Please clean
before reuse as follow:
• Clean the device with a soft, dry cloth with 70%
alcohol.
• Do not use petrol, thinners or similar solvent.
• Clean the cuff carefully with cloth soaked 70%
alcohol.
• The cuff must not be washed!
• Clean on the monitor and the arm cuff, and
then let it air dry.
Note: The device is a non-sterile medical device
and does not contain any sterile or degradable
component thus the device is not subject to the
shelf life requirements.
5
Troubleshooting
6
Specifications
Environmental
Operating
Storage
Temperature
5 to 40°C
-25 to 70°C
Relative humidity
(noncondensing)
10% to 95%
10% to 95%
Barometric
700 to 1060hPa
700 to 1060hPa
Protection
against electric
shock
Internally powered equipment
Degree
protection
against electrical
shock
Type BF
Electro-magnetic
compatibility
Group I, Class B
Degree of dust &
water resistance
IP22
Weight
15 g
Size
38×30×38
mm
Battery
3.7Vdc, Rechargeable Lithium-
polymer
Charge
requirement
5VDC, Max. 40mA
Charge time
2-3 hours
Battery life
12-16 hours for typical use
Wireless
Bluetooth 4.2 BLE
Oxygen level
range
70% to 100%
SpO2 Accuracy
(Arms)
80-100%:±2%, 70-79%:±3%
A functional tester or SpO2 simulator can be used to
determine the pulse rate accuracy.
Pulse Rate range
30 to 250 bpm
Pulse Rate
accuracy
±2 bpm or ±2%, whichever is
greater
Wavelength / Max
emission power
660nm/940nm, 0.8mW/1.2mW
SpO
2
data
averaging time
8s
SpO
2
data update
period
1s
Vibration source
low oxygen level;
high/low pulse rate
Recorded
parameters
Oxygen level, Pulse Rate
Record interval
4s
Data storage
4
records
, up to 10 hours for each
Frequency range
2.402
–
2.480 GHz
Max RF power
-10 dBm
Expected service
life
3 years
7
FCC Warnning:
Any Changes or modifications not expressly
approved by the party responsible for compliance
could void the user's authority to operate the
equipment.
Problem
Possible Cause
Possible Solution
Device
does not
turn on
or no
response
Battery may be low.
Charge battery and try
again.
Device might be
damaged.
Please contact your
local distributor.
Software exception
Keep device in
charging, touch the
key for 8 seconds.