background image

0,01

/

0.02

0.03

0,1

/

0.06

0.11

1

/

0.18

0.35

10

/

0.57

1.1

100

/

1.8

3.5

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)

can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power

rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

Rated maximum output

power of transmitter(W)

Separation distance according to frequency of transmitter(m)

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

Recommended separation distances between portable and mobile RF communications equipment

 and the Medical FS10C/FS20C PULSE OXIMETER

The FS10C/FS20C Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are

controlled. The customer or the user of the Medical FS10C/FS20C PULSE OXIMETER can help prevent electromagnetic

interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)

and the FS10C/FS20C Pulse Oximeter as recommended below, according to the maximum output power of the

communications equipment.

Conducted RF
IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

Portable and mobile RF communications equipment should be used no closer
to any part of the FS10C/FS2C Pulse Oximeter, including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.

Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths

from fixed

RF transmitters,

as determined

by an

electromagneti

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.

Compliance level

Radiated RF

IEC 61000-4-3

Guidance and manufacturer’s declaration – electromagnetic immunity

The FS10C/FS20C Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the
FS10C/FS20C Pulse Oximeter should ensure that it is used in such an environment.

Immunity test

IEC 60601 test level

N/A

3 V/m

3 V/m

80 MHz to 2,5 GHz

                                     80MHz to 800 MHz

 
                                     800MHz to 2.5 GHz

Electromagnetic environment 

– 

guidance

Trouble

Possible Reason

Solution

The batteries are completely exhausted

Please replace the batteries

An incorrect battery installation.

Verify and correct the batteries installation

The oximeter breakdown

Please contact local service

The batteries are completely exhausted

Please replace batteries

The luminescent or  photoelectric window is

sheltered by some object.

Check the luminescent and  photoelectric window.

Excessive movement

Stop moving finger,hand and body

The finger is not placed inside deep enough.

Place the finger properly and try again.

Finger size is not within the recommended range

Change another finger

Excessive ambient light

Avoid the excessive light

Pulse rate value of the cyclical fluctuations.

The measurement is normal, and the patient is

arrhythmia.

The finger is not properly positioned

Place the finger properly and try again.

The patient's SpO2 is too low to be detected

Try again, Go to a hosptial for a diagnosis if you are

sure the device works all right

Press the button again to reactivate the oximeter

The device will turn into sleep mode automatically

if there is no signal in 8 seconds

The oximeter can't turn to

measure mode

The display is off suddenly.

The Spo2 and Pulse Rate

display unstable

The Spo2 and PR are not

displayed normally

Electrostatic

discharge (ESD)

IEC 61000-4-2

NOTE U T is the a.c. mains voltage prior to application of the test level.

Power frequency

(50/60 Hz)

magnetic field

Guidance and manufacturer”s declaration ‘ electromagnetic immunity

The FS10C/FS20C Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the

user of the FS10C/FS20C Pulse Oximeter should ensure that it is used in such an environment.

Immunity test

Floors should be wood, concrete or ceramic tile. If floors are

covered with synthetic material, the relative humidity should be at

least 30 %.

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

IEC 60601

test level

Compliance

level

Electromagnetic environment ‘guidance

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic

of a typical location in a typical commercial or hospital environment.

Emissions test

Compliance

RF emissions

CISPR 11

Group 1

RF emissions

CISPR 11

Class B

Harmonic emissions

IEC 61000-3-2

N/A

Voltage fluctuations

flicker emissions

IEC 61000-3-3

N/A

The FS10C/FS20C Pulse Oximeter is suitable for use in all establishments, including

domestic establishments and those directly connected to the public low-voltage power

supply network that supplies buildings used for domestic purposes.

Guidance and manufacturer”s declaration ‘ electromagnetic emissions

The FS10C/FS20C Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the

user of the FS10C/FS20C Pulse Oximeter should ensure that it is used in such an environment.

Electromagnetic environment 

ẅ 

guidance

The FS10C/FS20C Pulse Oximeter uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to cause any interference in

nearby electronic equipment.

Bias Analysis

SpO2-FS20C

Pulse Oximeter

70 – 80 (%)

80 – 90 (%)

90 – 100 (%)

70 – 100 (%)

Mean Bias (B

s

)

1.94

1.45

0.89

1.4

Precision (S

res

)

2

1.55

0.98

1.53

Accuracy (A

rms

)

1.98

1.53

0.96

1.52

SaO2-Radiometer ABL800 FLEX-CO-Oximeter

7  Maintenance,Cleaning,Disinfection

8  Cleaning / Disinfection

One user manual
One certificate card

File Ver: A