11
6
Specifications
Classifications
EC Directive
MDD, 93/42/EEC
R&TTE, 1999/5/EC
ROHS 2.0, 2011/65/EU
Degree
protection
against
electrical
shock
Type BF
Environmental
Item
Operating
Storage
Temperature
5 to 40°C
-25 to 70°C
Relative
humidity
(noncondensing)
10% to 95%
10% to 95%
Barometric
700 to 1060
hPa
700 to 1060 hPa
Degree of dust &
water resistance
IP22
Physical
Weight
12 g (main unit)
Display
OLED
Wireless
Bluetooth 4.0
BLE
Vibrator
Built in
Power Supply
Charge input:
DC 5V ±10%
Battery type
Rechargeable lithium-polymer battery
Charge time
2~3 hours
SpO
2
Standards
Meet standards of ISO 80601-2-61
Measurement accuracy verification: The SpO
2
accuracy
has been verified in human experiments by comparing with
arterial blood sample reference measured with a CO-
oximeter. The pulse rate accuracy has been verified by
Emulator. Pulse oximeter measurement are statistically
distributed and about two-thirds of the measurements are
expected to come within the specified accuracy range
compared to CO-oximeter measurements.
SpO
2
range
70% to 100%