Section 1: Using the ERM – 8010 Thermal Printer
Page 7 of 7
The following summarises the procedure:
Turn printer ON Press and hold MODE button down until Error Lamp has flashed
at least 5 times. On release of the MODE button the Data lamp will flash a
number of times:
Data Lamp flashes once:
Error lamp flashes once
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes twice:
Error lamp flashes 7 times
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 3 times:
Error lamp flashes twice
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 4 times:
Error lamp flashes once
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 5 times:
Error lamp flashes 3 times
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 6 times:
Error lamp flashes once or twice
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 7 times:
Error lamp flashes once
If correct, Press FEED button otherwise press MODE button until correct.
Data Lamp flashes 8 times:
Error lamp flashes once – procedure
ended.
To save the changes made it is necessary to now press and hold the MODE and
the FEED buttons simultaneously. Hold them down for 5 seconds before
releasing them. On release the printer will print a summary of the settings for
confirmation. Confirm that they match the required settings specified on page 6.
Using the printer with the power supply connected
In some circumstances, using the power supply for the printer whilst performing
an ECG reading for example, may result in mains interference on the printed
trace. This may not always be apparent and will vary from location to location. If
interference is noted then disconnect the charger whilst printing ECG traces.