V1407 User Manual
48
ECG Out Function (Optional)
The V1407 provides ECG analog/pulse wave output from a rear stereo connector. It can be used with
defibrillator sync input or other medical devices.
How to get ECG Output
1
Connect the cable to the rear radio connector of the
V1407.
2
The blue waveform indicates an ECG analog signal. It can
be measured at the analog pin with a 1mV/mV unit. The
pink waveform indicates an ECG pulse detect signal. It
can be measured at a defibrillator sink pin with a 0–5V
unit.
3
The ECG waveform out cable recommended by the manufacturer is described below.
ECG Analog Signal
Defibrillator Sync Signal
ECG Output Port