Scale Factor adjustment
After you have calibrated your meter, you can use the SCALE feature to make fine adjust-
ments to calibration, without affecting the calibration itself.
Examples
1.Changing weight units of measure from kg to pounds
You could also use the SCALE to convert your readout from kg to pounds, without affecting
the calibration. Simply set SCALE = 2.205 and your meter which was calibrated in kg will
now read in pounds.
2. Correcting for gravitational variance
Your weighing system was calibrated where gravitational acceleration = 9.812m/s
2
(Lon-
don) You then move the system to Bankok where gravitational acceleration is reduced to
9.782m/s
2
You can correct for this difference by setting Scale = 9.812 / 9.782 = 1.003, so that a given
mass in Bangkok will show the same weight as it did in London. Set Offset = 0.0000
See
http://en.wikipedia.org/wiki/Earth%27s_gravity
Lockout Switch must be OFF
Press 3 seconds
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Press repeatedly until you see
SCALE
, followed by the existing scale
factor.
(Default =
001.000
)
1
2
3
4
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Use DIGIT button to select each digit
in turn, UP or DOWN buttons to
increase or decrease each digit’s
value. Press OK when done.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Done!
!! This feature is available in Advanced Mode only !!
You may want to adjust an offset value also, see separate OFFSET page for this feature.
Press to accept