Last update: 2022/07/20 23:32
slice:qt
https://www.vescent.com/manuals/doku.php?id=slice:qt
https://www.vescent.com/manuals/
Printed on 2022/07/21 23:14
Fig. 30: Slow Servo enabled on channel 1
The setpoint temperature of the slow servo channel can still be adjusted, by touching the Setpoint
[°C] field. To adjust the slow servo settings, press and hold on the blue Setpoint [°C] Field. This will
bring up the slow servo menu as shown in (
).
Fig. 31: Slow Servo menu showing adjustable parameters
This can be done either from the home screen or the single channel summary menu. Gain, polarity,
and setpoint can also be adjusted from the I/O menu by pressing the “gear” icon next to the A or B
input where the slow servo is enabled.
Choice
Comment
Off
Value not read
External Set Point
Externally control the set point
temperature
External Temp
Input
Alternate input path for plant
temperature
External Error
Input
Alternate input path for error signal
(bypasses wheatstone bridge)
Feed Forward
Sums input value with loop filter output.
Slow Servo Input
Input is routed through an integrator and
then used to control the set point
Tab. 7: Front-panel input signals