Vernier STS-BTA Manual Download Page 1

Surface Temperature
Sensor

(Order Code STS-BTA)

The Surface Temperature Sensor is designed for use in
situations in which low thermal mass or flexibility is
required. Special features include an exposed thermistor
that results in an extremely rapid response time.
Typical uses for the Surface Temperature Sensor include the following:

Skin temperature measurements

Human respiration studies

Specific heat experiments

Heat transfer experiments

Friction and energy studies

Note:

Vernier products are designed for educational use. Our products are not

designed nor are they recommended for any industrial, medical, or commercial
process such as life support, patient diagnosis, control of a manufacturing
process, or industrial testing of any kind.

What's Included

Vernier Surface Temperature Sensor

Compatible Software and Interfaces

See

www.vernier.com/manuals/sts-bta

for a list of interfaces and software compatible

with the Surface Temperature Sensor.

Getting Started

1. Connect the sensor to the interface (LabQuest Mini, LabQuest 2, etc.).
2. Start the appropriate data-collection software (Logger

Pro

, Logger Lite,

LabQuest App) if not already running, and choose New from File menu.

The software will identify the sensor and load a default data-collection setup. You
are now ready to continue your experiment.
If you are collecting data using a Chromebook™, mobile device such as iPad

®

or

Android™ tablet, or a Vernier wireless sensor or interface, please see the following
link for up-to-date connection information:

www.vernier.com/start/sts-bta

Using the Product

Connect the sensor following the steps in the Getting Started section of this user
manual.
The Surface Temperature Sensor is for use in air and water only. For temperature
measurements in harsher environments that require a more durable probe, we
recommend our Stainless Steel Temperature Probe (order code TMP-BTA).

Videos

View videos related to this product at

www.vernier.com/sts-bta

Calibration

It should not be necessary to perform a calibration when using this sensor. It is
calibrated before it ships. However, you can calibrate the sensor using Logger

Pro

3.3. or newer.

Note:

Calibration can only be done on computers running

Logger 

Pro

. For more information, see

www.vernier.com/til/1310

Specifications

Temperature range

–25 to 125°C (–13 to 275°F)

Maximum temperature that the
sensor can tolerate without damage

150°C

13-bit resolution

0.04°C (–25 to 0°C)
0.02°C (0 to 40°C)
0.05°C (40 to 100°C)
0.13°C (100 to 125°C)
0.08°C (–25 to 0°C)

12-bit resolution

0.08°C (–25 to 0°C)
0.03°C (0 to 40°C)
0.1°C (40 to 100°C)
0.25°C (100 to 125°C)

10-bit resolution

0.32°C (–25 to 0°C)
0.12°C (0 to 40°C)
0.4°C (40 to 100°C)
1.0°C (100 to 125°C)

Temperature sensor

20 k

NTC Thermistor

Accuracy

±0.2°C at 0°C, ±0.5°C at 100°C

Response time
(time for 90% change in reading)

50 seconds (in still air)
20 seconds (in moving air)

Probe dimensions: probe length
(handle plus body)

15.5 cm

How the Sensor Works

This probe uses the 20 k

NTC Thermistor, which is a variable resistor. When the

temperature increases, the resistance decreases non-linearly. The best-fit
approximation to this nonlinear characteristic is the Steinhart-Hart equation. At
25°C, the resistance is approximately 4.3% per °C. The interface measures the
resistance value,

R

, at a particular temperature and converts the resistance using the

2

Reviews: