Vernier Go Direct GDX-ETOH Manual Download Page 4

4

your Go Direct Ethanol Vapor, contact Vernier Technical Support at
[email protected] or call 888-837-6437. Support specialists will work with
you to determine if the unit needs to be sent in for repair. At that time, a Return
Merchandise Authorization (RMA) number will be issued and instructions will
be communicated on how to return the unit for repair.

Accessories/Replacements

Item

Order Code

Micro USB Cable

CB-USB-MICRO

USB-C to Micro USB Cable

CB-USB-C-MICRO

Go Direct 650 mAh Replacement Battery

GDX-BAT-650

Ethanol Cap Assemblies (package of 3)

ETH-CAPS

#6 Stopper

ETH-STOP

Plumber's Tape

ETH-TAPE

250 mL Nalgene Bottle w/Lid

CO2-BTL

BioChamber 2000

BC-2000

Electrode Support

ESUP

Stir Station

STIR

Warranty

Vernier warrants this product to be free from defects in materials and
workmanship for a period of five years from the date of shipment to the
customer. This warranty does not cover damage to the product caused by abuse
or improper use. This warranty covers educational institutions only. Consumable
items, such as Plumber's Tape, are excluded from the warranty.

Disposal

When disposing of this electronic product, do not treat it as household waste. Its
disposal is subject to regulations that vary by country and region. This item
should be given to an applicable collection point for the recycling of electrical
and electronic equipment. By ensuring that this product is disposed of correctly,
you help prevent potential negative consequences on human health or on the
environment. The recycling of materials will help to conserve natural resources.
For more detailed information about recycling this product, contact your local
city office or your disposal service.

Battery recycling information is available at

www.call2recycle.org

Do not puncture or expose the battery to excessive heat or flame.

The symbol, shown here, indicates that this product must not be disposed of

in a standard waste container.

Federal Communication Commission Interference Statement

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If

If any liquid gets into the device, immediately power the unit down (press and
hold the power button for more than three seconds). Disconnect the sensor and
charging cable, and remove the battery. Allow the device to dry thoroughly
before attempting to use the device again. Do not attempt to dry using an
external heat source.

Note:

Damage due to exposure to liquids is not covered under warranty.

How the Sensor Works

Go Direct Ethanol Vapor uses a metal oxide semiconductor to detect ethanol
concentrations in the air. Ethanol vapor reacts with the metal oxide and is
consumed in a reaction that reduces the resistance of the sensing element. A
circuit converts the change in resistance into an appropriate ethanol
concentration. The sensing element is very sensitive to ethanol vapor and also
responds to a wide range of volatile organic compounds. It also responds to
water vapor and carbon dioxide gas but to a lesser degree.

Troubleshooting

l

Very important:

Do not place the sensor into any liquid. The sensor is

designed to measure ethanol vapor above an aqueous sample.

l

For best results, allow the sensor to warm up for 5 minutes before taking
measurements.

l

Built-in temperature, pressure, and humidity compensation are not provided.
For best results, calibrate before use and use at room temperature.

l

The recommended sampling rate is 1 sample per second or slower for this
sensor. This allows the sensor to average a set of measurements before
reporting a value. Gas must diffuse to the sensing element to detect any
changes in concentration. Because diffusion of gases is a slow process, there
can be a delay in response time.

l

When using a fermentation chamber, use a clamp or an electrode support to
adjust the position of the sensor tip by moving the shaft up or down as
necessary.

l

The 250 mL Nalgene Bottle and #6 stopper that are included can be used as
a fermentation chamber. Wrap the stopper around the sensor shaft, rather
than sliding it on or off. Place the probe with the stopper into the bottle, and
then gently twist the stopper. Verify that the sensor tip is not immersed in
fluid. For best results, place the chamber on a stir plate and use a magnetic
stir bar to keep the sample moving.

l

To collect data in a controlled environment and also use our one of our CO

2

gas sensors, we recommend purchasing the BioChamber 2000 (2 L capacity).
This BioChamber has a second, grommetted opening so that you can insert
two probes at once.

For more troubleshooting and FAQs, see

www.vernier.com/til/4303

Repair Information

If you have followed the troubleshooting steps and are still having trouble with

Reviews: