background image

2

3. Fill a cuvette about 3/4 full with distilled water (or the solvent being used

in the experiment) to serve as the blank. After the Spectrophotometer has
warmed up, place the blank cuvette in the Spectrophotometer. Align the
cuvette so the clear side of the cuvette is facing the light source. Click or tap
Finish Calibration.

4. You are now ready to collect data. Fill a cuvette about 3/4 full of a sample

of the solution to be tested. Place the sample in the Spectrophotometer and
click Collect. Click Stop to end data collection. The spectrum is
automatically stored.

Measurement

vs.

Concentration (Beer's law)

1. Select Measurement

vs.

Concentration.

2. The calibration dialog box will appear. The warm-up timer starts counting

when the spectrometer connects to the platform. As a result, the countdown
may appear to vary. The minimum warm-up time is 90 seconds. For best
results, we suggest a warm-up time of several minutes.

3. Fill a cuvette about 3/4 full with distilled water (or the solvent being used

in the experiment) to serve as the blank. After the Spectrophotometer has
warmed up, place the blank cuvette in the Spectrophotometer. Align the
cuvette so the clear side of the cuvette is facing the light source. Click or tap
Finish Calibration.

4. Follow the instructions in the Choose a Wavelength dialog box or simply

type in the wavelength you wish to measure. Select Done.

5. Click Collect. Your first sample should still be in the Spectrophotometer.

After the reading stabilizes, click Keep. Enter the concentration of the
sample and click or tap Keep Point.

6. Place your second sample in the cuvette slot. After the reading stabilizes,

click Keep. Enter the concentration of the sample and click or tap Keep
Point.

7. Repeat Step 6 for the remaining samples. When finished, click Stop to end

data collection. The data is automatically stored.

8. To see the best fit line equation for the standard solutions, click Graph

Tools, select Apply Curve Fit, and choose Linear. Click or tap Apply.

9. If doing Beer's law to determine the concentration of an unknown, place the

unknown sample in the cuvette holder. Click or tap Graph Tools and enable
Interpolate. Click or tap along the line until you find the concentration
value that matches your unknown's measurement.

10. Save or export your data from the File menu.

Measurement

vs.

Time (Kinetics)

1. Select Measurement

vs.

Time.

2. The calibration dialog box will appear. The warm-up timer starts counting

when the Spectrophotometer connects to the platform. As a result, the
countdown may appear to vary. The minimum warm-up time is 90 seconds.
For best results, we suggest a warm-up time of several minutes.

3. Fill a cuvette about 3/4 full with distilled water (or the solvent being used

in the experiment) to serve as the blank. After the Spectrophotometer has

Connecting via Bluetooth

Ready to connect

Press Power button once. Blue LED next to
Bluetooth icon flashes when sensor is ready to
connect.

Connected

Blue LED next to Bluetooth icon is solid when
sensor is connected via Bluetooth wireless tech-
nology.

Connecting via USB

Connected and charging

Orange LED next to battery icon is solid when
sensor is connected to Spectral Analysis via
USB. LED next to Bluetooth icon is off.
LED next to power is solid green. LED next to
USB icon is solid green.

Connected, fully charged

LED next to battery icon is off when sensor is
connected to Spectral Analysis via USB and
fully charged. LED next to Bluetooth icon is off.

Charging via USB, con-
nected via Bluetooth

This combination is not permitted.

Charging via AC power,
connected via Bluetooth

This combination is not recommended.

Using the Product with Spectral Analysis 4

Connect the sensor following the steps in the Getting Started section of this user
manual.

Collect Data with Spectral Analysis 4

The three options for Experiment Types are:

1. Measurement

vs.

Wavelength—collect a full spectrum.

2. Measurement

vs.

Concentration—conduct a Beer's law experiment.

3. Measurement

vs.

Time—collect time-based data for a kinetics experiment.

By default, Absorbance is selected. If you want to measure %Transmittance, use
the toggle switch.

Note:

Visit

www.vernier.com/spectral-analysis

for status of

intensity and fluorescence support.

Measurement

vs.

Wavelength (Full Spectrum)

1. Select Measurement

vs.

Wavelength.

2. The calibration dialog box will appear. The warm-up timer starts counting

when the Spectrophotometer connects to the platform. As a result, the
countdown may appear to vary. The minimum warm-up time is 90 seconds.
For best results, we suggest a warm-up time of several minutes.

Summary of Contents for GDX-SVISPL

Page 1: ...r and fluorometer This spectrophotometer can be used in a wide range of introductory spectroscopy experiments for chemistry biology and physics Such experiments include determining the peak wavelength...

Page 2: ...you find the concentration value that matches your unknown s measurement 10 Save or export your data from the File menu Measurement vs Time Kinetics 1 Select Measurement vs Time 2 The calibration dia...

Page 3: ...sh to measure Select Done 5 The default data collection settings collect measurements every two seconds USB or three seconds BLE until the user manually stops data collection 6 Mix the reactants Trans...

Page 4: ...tion If the spectrum maxes out flat and wide peaks at a value of 1 increase the distance between the light source and the tip of the optical fiber cable or reduce the sample time see Change the Settin...

Page 5: ...Spectra Open the desired file You can use the mercury emissions graph to test fluorescent lighting for the presence of mercury Measure Fluorescence with Logger Pro You may use your Spectrophotometer...

Page 6: ...menu 4 On the Meter screen tap Change Units USB Spectrophotometer Intensity from the Sensors menu Intensity is a relative measure with a range of 0 1 Note The Spectrophotometer is not calibrated for...

Page 7: ...asure Fluorescence with LabQuest You may use your Spectrophotometer to measure the fluorescence spectrum of an aqueous sample such as chlorophyll riboflavin and fluorescein Fluorescence is the emissio...

Page 8: ...lectrique m me si cela r sulte un brouillage susceptible d en compromettre le fonctionnement Cet appareil num rique respecte les limites de bruits radio lectriques applicables aux appareils num riques...

Page 9: ...re our trademarks or registered trademarks in the United States All other marks not owned by us that appear herein are the property of their respective owners who may or may not be affiliated with con...

Reviews: