Vermason TB-7588 Installation, Operation And Maintenance Manual Download Page 2

DESCO EUROPE

 - 2A DUNHAMS LANE, LETCHWORTH, HERTFORDSHIRE, SG6 1BE, UK

Phone: +44 (0) 1462 672005 • E-mail: 

[email protected]

, Website: 

DescoEurope.com

TB-7588

 Page 2 of 7

© 2017 DESCO INDUSTRIES, INC.

Employee Owned

Packaging

222642

 Digital Surface Resistance Meter Kit

1  Digital Surface Resistance Meter

2  Shielded Test Leads, 1.5 m Length

2  2 kg Electrodes

2  AA Alkaline Batteries

1  Gator Clip

1  BNC / Banana Jack Adapter

1  Plastic Carrying Case

1  Certificate of Calibration

Figure 2.  Vermason 

222642

 Digital Surface Resistance 

Meter Kit

Features and Components

Figure 3.  Vermason 

222643

 Digital Surface Resistance 

Meter

222643

 Digital Surface Resistance Meter

1  Digital Surface Resistance Meter

2  AA Alkaline Batteries

1  Certificate of Calibration

FRONT VIEW

REAR VIEW

A

B

C

D

E

F

G

H

Figure 4.  Digital Surface Resistance Meter features and 

components

Summary of Contents for TB-7588

Page 1: ...arments ANSI ESD STM2 1 Garments Person footwear flooring system IEC 61340 4 5 A compliance verification plan shall be established to ensure the Organization s fulfilment of the technical requirements...

Page 2: ...m Length 2 2 kg Electrodes 2 AA Alkaline Batteries 1 Gator Clip 1 BNC Banana Jack Adapter 1 Plastic Carrying Case 1 Certificate of Calibration Figure 2 Vermason 222642 Digital Surface Resistance Mete...

Page 3: ...meter will cycle through the approximate 15 second electrification period that displays the temperature and relative humidity before the surface resistance mantissa is displayed When set to QUICK the...

Page 4: ...if preferred If the measurement is outside acceptable limits clean the surface with an ESD cleaner containing no insulative silicone such as Reztore Antistatic Surface Mat Cleaner and re test to dete...

Page 5: ...guide clause 4 3 3 Verification frequency The frequency of periodic testing is normally specified in corporate operating procedures However a common guide would be to conduct the measurements of EPA...

Page 6: ...xcept at 1011 and 1012 5 Thermometer accurate to 1 F Relative Humidity Meter accurate to 2 Test Leads 99 Isopropyl Alcohol and Cleaning Wipes Setup Test Area needs to be free of any high voltage trans...

Page 7: ...x 103 Temperature 24 C 1 7 C 75 F 3 F Relative Humidity 10 integers A 1 0 x 1012 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 1011 20 LED 11 Yell...

Reviews: