Vermason TB-7588 Installation, Operation And Maintenance Manual Download Page 1

DESCO EUROPE

 - 2A DUNHAMS LANE, LETCHWORTH, HERTFORDSHIRE, SG6 1BE, UK

Phone: +44 (0) 1462 672005 • E-mail: 

[email protected]

, Website: 

DescoEurope.com

TB-7588

 Page 1 of 7

© 2017 DESCO INDUSTRIES, INC.

Employee Owned

Digital Surface Resistance Meter Kit

Installation, Operation and Maintenance

Description

The Vermason 

222642

 Digital Surface Resistance Meter 

Kit is an instrument designed to measure resistance 

point-to-point (Rp-p) or surface to ground (Rg) in accordance 

with EN 61340-5-1 Electrostatics and its test method IEC 

61340-2-3.
Its test functions include:

•  Resistance measurement accuracy of ±10% 

(±20% for resistance values 1 x 10

11

 ohms and greater)

•  Resistance range of <1 x 10

3

 

ohms to >1 x 10

12

 

ohms

•  Under load voltages of 10 and 100 volts ±5%

•  Electrification period of approximately 15 seconds
The Digital Surface Resistance Meter also measures ambient 

temperature and relative humidity.
NOTE: Volume Resistance can be measured using one 2 kg 

Electrode and a flat conductive metal plate (not included).
The Digital Surface Resistance Meter Kit and its accessories 

are available in the following item numbers:

November 2017

Figure 1.  Vermason Digital Surface Resistance Meter 

Kit

The Surface Resistance Meter Kit (or its Meter) is referenced 

and designed to be used to make measurements in 

accordance with the test procedures in: 
•  Working surfaces, storage racks, trolleys and seating - 

IEC 61340-2-3 

•  Flooring - IEC 61340-4-1 
•  Garments - ANSI/ESD STM2.1 Garments 
•  Person/footwear / flooring system - IEC 61340-4-5
“A compliance verification plan shall be established to ensure 

the Organization’s fulfilment of the technical requirements 

of the plan. Process monitoring (measurements) shall be 

conducted in accordance with a compliance verification 

plan that identifies the technical requirements to be verified, 

the measurement limits and the frequency at which those 

verifications shall occur. The compliance verification plan 

must document the test methods and equipment used for 

process monitoring and measurements. If the organization 

uses test methods that differ from the standards referenced 

in this standard, the organization must be able to show that 

the results achieved correlate with the referenced standards.  

Compliance verification records shall be established and 

maintained to provide evidence of conformity to the technical 

requirements.

The test equipment selected shall be capable of making the 

measurements defined in the compliance verification plan.” 

(IEC 61340-5-1 Edition 1 2007-08 clause 5.2.3 Compliance 

verification plan)

Reference Literature

In addition to those noted above:

EN 61340-5-1 Electrostatics and PD CLC/TR 61340-5-2 User 

Guide Electrostatics
These documents can be obtained from from British 

Standards at 

www.bsigroup.com

.

Made in the

United States of America

Item

Description

222642

Digital Surface Resistance Meter Kit

222643

Digital Surface Resistance Meter

222002

Concentric Ring Probe

222003

Concentric Ring Probe Kit

222633

Shielded Test Leads

222634

2 kg Electrodes, 1 Pair

  TECHNICAL BULLETIN TB-7588

Summary of Contents for TB-7588

Page 1: ...arments ANSI ESD STM2 1 Garments Person footwear flooring system IEC 61340 4 5 A compliance verification plan shall be established to ensure the Organization s fulfilment of the technical requirements...

Page 2: ...m Length 2 2 kg Electrodes 2 AA Alkaline Batteries 1 Gator Clip 1 BNC Banana Jack Adapter 1 Plastic Carrying Case 1 Certificate of Calibration Figure 2 Vermason 222642 Digital Surface Resistance Mete...

Page 3: ...meter will cycle through the approximate 15 second electrification period that displays the temperature and relative humidity before the surface resistance mantissa is displayed When set to QUICK the...

Page 4: ...if preferred If the measurement is outside acceptable limits clean the surface with an ESD cleaner containing no insulative silicone such as Reztore Antistatic Surface Mat Cleaner and re test to dete...

Page 5: ...guide clause 4 3 3 Verification frequency The frequency of periodic testing is normally specified in corporate operating procedures However a common guide would be to conduct the measurements of EPA...

Page 6: ...xcept at 1011 and 1012 5 Thermometer accurate to 1 F Relative Humidity Meter accurate to 2 Test Leads 99 Isopropyl Alcohol and Cleaning Wipes Setup Test Area needs to be free of any high voltage trans...

Page 7: ...x 103 Temperature 24 C 1 7 C 75 F 3 F Relative Humidity 10 integers A 1 0 x 1012 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 1011 20 LED 11 Yell...