background image

GUIDELINES FOR TEMPLE MEASUREMENTS 

This thermometer is designed to measure the temperature at the forehead, in
proximity to the temple area between the end of the eyebrow and the hairline. 

Do not place the thermometer on scar tissue, skin disorder,
open sores or abrasions, or patients in trauma.

Consult your physician when using drug therapies that may
elevate skin temperature in the forehead.

Do not use skin exposed to direct sunlight,  fireplace heat,
air conditioner or fan air flow, or hot/cold compress
therapies. 

Wait at least one minute between measurements to
allow the unit sensor to readjust to room
temperature.

If the thermometer has been stored in an excessively cold or
warm environment, allow the thermometer to adjust to the
ambient room temperature before use. Performance and
operation of this unit may be compromised if operated
outside of the specified temperature and humidity
operating ranges.

OPERATING INSTRUCTIONS 

1.

Press and release the      POWER button to activate the
thermometer.

2.

The unit will beep and the thermometer will perform a
self-test, showing all of the symbols on the display, Fig 1.

3.

The last measurement taken will show on the display.  

4.

Gently place the probe in the center of the temple area
between the end of the eyebrow and the hairline.
Maintain firm contact with the skin, hold the
thermometer in place during the reading and do not
move the thermometer to ensure reliable results. 

5.

Do NOT press the power button again. The measurement
will start automatically once the probe is in place and will
take approximately 6-8 seconds. A running circle will
appear on the display indicating the reading is in
progress, Fig 2. 

6.

The unit will sound one long beep when the
measurement is complete and the reading will appear on the display, Fig 3. 

7.

To turn the unit off, simply press the      POWER button. The unit will automatically
shut-off after approximately one minute. 

NOTE

: If you take another measurement before 1 minute has elapsed, the device

will perform a 10 to 0 countdown to allow the device to reset for a new
measurement. 

DO NOT TRY TO MEASURE DURING THIS COUNTDOWN. If you attempt to measure
during the countdown time, the device will display the previous result, beep 8
times and turn itself off automatically.

CHANGING MODES - FAHRENHEIT OR CELSIUS OPERATION 

This unit defaults to a Fahrenheit measurement. To change modes:

1.

Press and hold the      POWER button for approximately 
5-seconds.

2.

The unit will beep, then the display will indicate the
self-test screen, last measurement and finally will
show only the current mode in the upper right
corner. 

IN

S

T

R

U

C

T

IO

N

S

 •

 E

N

G

L

IS

H

THANK YOU FOR PURCHASING A MINI TEMPLE THERMOMETER. PLEASE READ THESE
INSTRUCTIONS CAREFULLY TO ENSURE PROPER USE AND SAFE OPERATION OF YOUR
THERMOMETER. 

PLEASE READ AND SAVE THESE INSTRUCTIONS

This product is not intended to diagnose or treat any health problem or disease
and should not be used as a substitute for the consultation and advice of a
physician or other medical professional. 

Thermometers should only be used under the supervision of an adult. 

Before use, it is recommended that the patient avoid eating, drinking, smoking
or exercising for at least twenty minutes prior to taking a measurement. Never
walk, run or talk during a measurement.

NORMAL BODY TEMPERATURE AND FEVER 

Body temperature can vary from 97°F and 100.0°F and still be considered normal.
The medically accepted ‘normal’ body temperature is 98.6°F Body temperature is
commonly lower in the morning than at any point during the rest of an individual’s
waking hours. 

Temperature readings will vary based on the body location point of the reading.
Average ‘normal’ temperature ranges by measurement location site:
Forehead

97.0°F–100.0°F (36.1°C–37.8°C)

Oral

97.0°F–100.0°F (36.1°C–37.8°C)

Rectal

97.3°F–100.4°F (36.3°C–38.0°C)

Axillary

95.4°F–99.0°F (35.2°C–37.2°C)

A fever is defined as a body temperature that is elevated above the normal for that
person. It is important to determine what is normal for an individual before
determining if a fever is present. Tracking an individual’s temperature on a
consistent basis, at the same body site and at the same time everyday when the
person is well or healthy will help determine an individual’s normal temperature. 

QUICK START TIPS

1.

Press and release the      POWER button, wait 2-seconds.

2.

Hold unit firmly to temple until a long single beep is heard, 6-8 seconds.

3.

Results will be displayed on the screen. 

4.

Wait one minute before new measurement.

5.

To alternate between °F and °C, press and hold the      POWER button. 

3.

Continue to hold the      POWER button, the °F or °C will alternate to the desired
mode, either Celsius or Fahrenheit.  

4.

Release the      POWER button. Either proceed with your measurement, or press
and release      POWER to shut the unit off. 

MEMORY RECALL 

This unit includes a recall of the last measurement taken.
When the unit is OFF, press the      POWER button to turn the
unit on, and release. After the self-test, the unit will briefly
show the last reading on the display. 

REPLACING THE BATTERY

This thermometer includes 2 AAA 1.5V batteries. The low-battery symbol will appear
on the top center of the display when the battery needs to be changed. 

1.

Carefully slide the battery cap away from the unit.

2.

Replace the batteries, taking care to match the correct
polarity. 

3.

Carefully replace the battery cap. 

4.

Do not dispose of used batteries in the trash. Consult your
local recycling center for the proper way to dispose of
batteries in your area. Keep used batteries away from
children, pets and excessive heat.

  

TROUBLESHOOTING / IDENTIFYING DISPLAY SYMBOLS 

The thermometer has been designed to display feedback-warnings to ensure proper
usage and accurate temperature measurement. When an error appears, the device
will "beep" 5 times, and then the thermometer will turn off automatically. In the
event of a malfunction or incorrect temperature measurement, the following warn-
ings will be displayed:

PRODUCT SPECIFICATIONS

MEASUREMENT TIME:

6-8 seconds

MEASUREMENT SITE:

Temple area of forehead

MEASUREMENT RANGE:

95.0°F – 107.6°F (35° – 42°C)

OPERATING TEMPERATURE:

60.8°F – 104.0°F (16°C – 40°C); 
Humidity 15-95% (non condensing)

STORAGE TEMPERATURE:

-4.0°F – 122.0°F (-20°C – 50°C);

Humidity 15-95% (non condensing)

MEASUREMENT SCALE:

Fahrenheit and Celsius

MEMORY:

Recall of last measurement taken

BATTERY:

2 AAA 1.5V batteries 

BATTERY LIFE:

Approximately 24 months; 
minimum of 1000 measurements

AUTOMATIC SHUT-OFF:

Approximately after 1 minute of non-use

ACCURACY:

complies with ASTM E 1112-00 standards

DISPLAY:

Liquid crystal display

DISPLAY RESOLUTION:

LCD 0.63” x 0.44”

DIMENSIONS:

4.7” X 1.17” X 1”

WEIGHT:

1.3 ounces with batteries

Specifications are subject to change without notice.

  

ONE-YEAR LIMITED WARRANTY

Congratulations on your purchase of a Temple Touch Mini Thermometer.  Your Temple Touch Mini
Thermometer is covered by the following limited warranty commencing upon the date of purchase, and
subject to the following terms and conditions:

The Warrantor warrants that its Temple Touch Mini Thermometer will be free from defects in materials and
workmanship under normal consumer usage for a period of one year for the original purchaser of the Mini
Temple Thermometer. 

Periodic maintenance, repair and replacement of parts due to normal wear and tear are excluded from
coverage. Defects or damage that result from: (a) improper operation, storage, misuse or abuse, accident
or neglect, such as physical damage (cracks, scratches, etc.) to the surface of the product resulting from
misuse; (b) contact with liquid, water, rain, extreme humidity or heavy perspiration, sand, dirt or the like,
extreme heat, or food; (c) use of the Temple Touch Mini Thermometer for commercial purposes or
subjecting the Temple Touch Mini Thermometer to abnormal usage or conditions; or (d) other acts which
are not the fault of the Warrantor, are excluded from coverage. This warranty does not cover batteries or
other power sources that may be provided with, or used with the Temple Touch Mini Thermometer.

If the Digital Thermometer fails to conform to this limited warranty, return the Temple Touch Mini
Thermometer to: 

Attn: Repair Department, 1175 Lakeside Drive, Gurnee, IL, 60031.

When

returning a product, please also include: (i) a copy of your receipt, bill of sale or other comparable proof of
purchase; (ii) a written description of the problem; and (iii)  your name, address and telephone number.
Carefully package the product to avoid any damage that may occur while in transit; shipping insurance
with returned receipt is recommended. At our option, the warrantor will repair or replace the unit found
to be defective in materials or workmanship under normal consumer usage. The purchaser will be notified
of any additional repairs required prior to completing the repair, and will be responsible for parts charges,
if any, and repair charges not covered by this limited warranty.   

EXCEPT AS PROVIDED FOR IN THIS LIMITED WARRANTY, ALL EXPRESS AND IMPLIED WARRANTIES AND
CONDITIONS ARE DISCLAIMED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.  THE REPAIR OR REPLACEMENT AS
PROVIDED UNDER THIS LIMITED WARRANTY IS THE EXCLUSIVE REMEDY OF THE CONSUMER, AND IS
PROVIDED IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. IN NO EVENT SHALL THE WARRANTOR
BE LIABLE, WHETHER IN CONTRACT OR TORT (INCLUDING NEGLIGENCE) FOR DAMAGES IN EXCESS OF THE
PURCHASE PRICE OF THE PRODUCT, OR FOR ANY INDIRECT, INCIDENTAL, SPECIAL OR CONSEQUENTIAL
DAMAGES OF ANY KIND, OR FOR DAMAGES TO, OR LOSS OF, OTHER PROPERTY OR EQUIPMENT OR PERSONAL
INJURIES TO THE FULL EXTENT THESE DAMAGES MAY BE DISCLAIMED BY LAW.

Some states and jurisdictions do not allow the limitation or exclusion of incidental or
consequential damages, or limitation on the length of an implied warranty, so the above
limitations or exclusions may not apply to you. This warranty gives you specific legal rights,
and you may also have other rights that vary from state to state or from one jurisdiction to
another.

Toll-Free Customer Care Help Line: 1-866-326-1313

Monday – Friday 8:30 a.m. – 4:30 p.m. CST

Correct Placement

Incorrect Placement

Fig 1

Fig 2

Fig 3

ERROR 

PROBLEM/CAUSE

SOLUTION

The measured temperature is
above 107.6°F (42°C) 

The measured temperature is
below 95°F (35°C)

The thermometer is too cold or
too hot to operate

Temperature did not register 
because of poor contact, patient
movement  or poor placement

Unit malfunctioned

Low battery - Battery icon 
displays continuously

Drained battery - Battery icon is
the only icon the display and
unit turns off

1. Turn the unit OFF and then ON again 
2. Retake temperature
3. If still reads HI, this indicates that the pa-
tient has a high fever; seek medical attention

1. Turn the unit OFF and then ON again 
2. Retake temperature
3. If still reads LO, this indicates that the pa-
tient has a low body temperatur; seek med-
ical attention 

1. Verify that the room temperature within
product range 60.8°F-104°F (16° C - 40° C)  
2. Retake temperature 

Turn the unit OFF, wait 2 minutes, turn it ON
again and retake temperature 

Do not use the thermometer – contact
Customer Care 

Battery is low; enough power remains for
approximately 10 measurements

Battery is completely expired and must be
replaced

PROBE 
SENSOR

DIGITAL 
DISPLAY

POWER 
BUTTON

BATTERY 
COMPARTMENT

The Temple Touch  Thermometer works by
measuring the heat flow from the blood
vessels to the skin. 
That measurement is then converted to a
comparable body temperature. 

Bloo

d Ve

ssel

s

Hea

t/En

ergy

 Flo

w

Skin

Temple Touch 
Mini Thermometer

Model 09-330/09-331

Manufactured by: 
Medisim Ltd., Israel

Distributed by:
Veridian Healthcare
1175 Lakeside Drive
Gurnee, IL 60031
www.veridianhealthcare.com

Made in China

#93-1079  03/14
©2014 Veridian Healthcare, LLC

Protected by US patents

# 7,597,668 B2

Reviews: