2.0 Installation
Products without a factory fitted supply cable require a permanent connection to a fused
and isolated supply. The product's electrical access panel should be temporarily
removed, and connections made to the internal terminals.
If the product is to be connected by line plug. The plug should be within reach of the
operator and should be easy to remove.
When connecting the product to an isolating switch ensure that both conductors (single
phase) or on all live conductors (three phase), and should be within reach of the
operator.
The supply MUST incorporate an earth (ground).
Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase L
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
2- or
3-phase
L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3 (
except 2-phase
)
N
Light Blue
to neutral (
except delta
)
PE
Green/ Yellow
to earth (
ground
)
2.6
Voltage and Power Limit
When first starting up the product, check the setting of the OP.Hi
parameter (see the controller instructions) and compare it with the data
in section 10.0 towards the back of this manual.
If the power limit setting is incorrect for the voltage of the electrical supply, change it.
This must be done immediately if the setting of OP.Hi is higher than it should be.