5.0 Operation
5.5
Operator Safety
The ceramic materials used in the product manufacture become
electrically conductive to some extent at high temperatures. DO NOT use
any conductive tools within the product without isolating it. If a metal
work tube is used, it must be earthed (grounded).
Switch off the heater switch whenever loading or unloading the product.
The elements are isolated when the heater switch is OFF. This switch
cuts both sides of the circuit via a contactor.
5.6
Pressure
Work tubes are not able to accept high internal pressure. When gas seals or similar
fittings are in use, the gas pressure should be restricted to a maximum of 0.2 bar (3
psi). A pressure of approximately half of that should normally be sufficient to achieve
the desired flow rate. The operator must ensure that the exhaust path from the tube is
not blocked, so that excess pressure does not occur.
A suitably regulated gas supply should always be used.
It is recommended that a pressure relief system should be used to avoid an over
pressurisation of the work tube.
Please note: A product should not be heated up if any valves that have been fitted are
closed to create a sealed volume. A sealed work tube should not be heated from cold
due to the pressure increase caused by the trapped air or gas expanding during the
heating process.
26
Summary of Contents for CARBOLITE GERO CFM 14/2 AUX
Page 4: ...User Power Setting Adjustments 40 11 0 Specifications 41 11 1 Environment 41 4 ...
Page 6: ...1 0 Symbols and Warnings Caution Double Pole Neutral Fusing 6 ...
Page 20: ...3 0 2416 Controller 3 6 Navigation Diagram 20 ...
Page 37: ...9 0 Wiring Diagrams 9 1 WB 12 10 AUX 9 2 WB 12 11 AUX 37 9 0 Wiring Diagrams ...
Page 38: ...9 0 Wiring Diagrams 38 ...
Page 42: ......
Page 43: ...Notes Service Record Engineer Name Date Record of Work ...