4.0
Operation
4.1
Operating Cycle
This product is fitted with an instrument switch which cuts off power to the control
circuit.
Connect the product to the electrical supply.
Turn on the instrument switch to activate the temperature controllers. The controllers
illuminate and go through a short test cycle.
Over-Temperature option only
. If the hydraulic thermostat over-temperature option
is fitted, set the rotary dial to the desired protection temperature.
Over-Temperature option only
. If the digital over-temperature option has not yet
been set as required, set and activate it according to the over-temperature controller
instructions.
The product will heat up according to the controller setpoint or program, unless a time
switch is fitted and switched off.
Over-Temperature option only.
If the over-temperature circuit has tripped, an
indicator on the over-temperature controller flashes and the heating elements are
isolated. Find and correct the cause before resetting the over-temperature controller
according to the instructions supplied.
4.2
Over-Temperature Control (if fitted)
The over-temperature controller should typically be set at 15 °C above the main
controller. If an over-temperature condition occurs, check the main controller is
functioning correctly.
An over-temperature condition cuts off power to the heating elements. A light in the
over-temperature controller flashes. To reset this, refer to the over-temperature
control section of this manual.
4.3
Explosive Vapours
Unless your product includes the stoving and curing option, this
model is not suitable for drying or heat treatment applications where
vapours are released that are combustible or which can form
explosive mixtures with air. Carbolite Gero manufactures other
products suitable for these applications.
4.4
Coal Drying Application
The CDLT is designed to operate at typically 50°C to dry and test coal samples in
accordance with international standards.
The oven has air inlet and outlet vents designed to ensure a constant flow of fresh air
to dry the samples and carry moisture away.