Operations & Maintenance Manual: Maintenance & Troubleshooting
MAINTENANCE & TROUBLESHOOTING
REGULAR INSPECTIONS
Prior to each use, inspect the device for cracks, abrasions, gouging, evidence of impact, or other damage.
Cracks that allow the ingress of fluid may affect the safety and performance of the device.
If you see any physical damage or cracks in the device, discontinue use immediately and contact
Verathon
®
Customer Care or your local representative.
IMPORTANT
CALIBRATING THE DEVICE
You must periodically calibrate your device to ensure that it is providing accurate results. The required
frequency of calibration depends on your Total Reliability
SM
Plan. Calibrating the device on a regular basis
ensures accurate and proper alignment of the device’s internal coordinate system.
If you have the calibration kit and Scan Point
®
with QuickPrint, you can easily and quickly calibrate your own
device. You may also send your device to an authorized Verathon Service Center.
If you are not using Scan Point with QuickPrint, you must send your device in to an authorized Verathon
Service Center for calibration.
To contact Verathon Customer Care about calibration, please visit
PROCEdURE 1. CALIBRATE THE dEVICE
Calibrating your device ensures that it is providing accurate results, and maintains proper alignment of
the device’s internal coordinate system. If the device is not calibrated by the prescribed date, the device is
disabled and must be calibrated before it can be used again.
1. Place the calibration tank on a flat, non‑reflective surface, and then remove the lid.
2. Pour clean, room‑temperature water into the tank base, filling to the indicator mark. Ensure that there is
a minimal amount of bubbles in the water.
Note: The tank may need to sit for 24 hours until the water has degassed.