23
Operations & Maintenance Manual: Setting Up
0900‑4868 REV‑08
ATTACH THE CABLE & BRONCHOSCOPE
The GlideScope Video Monitor QuickConnect Cable attaches the bronchoscope to the monitor, supplying
power to the bronchoscope and transmitting video data from the camera to the monitor.
It is recommended that you leave the sterile, single‑use bronchoscope in the packaging while connecting
it, and that you do not remove the bronchoscope until you are ready to insert it. This helps ensure that the
bronchoscope remains as clean as possible.
1. Ensure that the video monitor is turned off.
2. Align the arrow on the GlideScope Video Monitor QuickConnect Cable and the arrow on the video cable
port.
3. Insert the cable connector into the port. You will hear a click when the cable is successfully connected.
4. Remove the bronchoscope and the introducer from their packets.
5. Detach the protective cover from the cable connector on the bronchoscope. Discard the cover after you
remove it.
6. Carefully slide the protective sleeve off the insertion tube of the bronchoscope. Discard the sleeve after
you remove it.
7. Inspect the bronchoscope to ensure that it is functional.