To configure VOCSN and begin providing High Flow therapy:
1
Select the
Therapy
tab on the touchscreen, then
Ventilation
.
To configure therapy, you must select an inactive
preset.
Note: If there is only one Preset visible on the screen, first press
Settings & Alarms
and use the
Preset Enable
control to enable a second Preset.
2
Press
Settings & Alarms
to configure the Preset you intend to use to deliver therapy. To change a control
setting, highlight the relevant control and then select
Edit
. Enter the new setting value, and then select
ACCEPT
.
Configure the inactive Preset as follows:
Control
Setting
Circuit Type
Valveless
Mode
SIMV-Pressure or SIMV-Volume
High Flow
On
Patient Type
<desired setting>
Flow
<desired setting>
3
Press
Exit
to return to the
Ventilation
screen, and then begin using the Preset you configured for High Flow. Press
the Preset to highlight it, and then press
START
beneath it.
4
Run a Pre-Use Test. If not prompted to begin one automatically, press the
Menu
tab, and then the
PRE-USE
TEST
button on the touchscreen.
Carefully follow the on-screen instructions.
Ensure the Pre-Use Test passes
successfully.
5
Next, begin
Oxygen
therapy. Select the
Therapy
tab on the touchscreen, then
Oxygen
. Select the Preset you
would like to configure, and then press
Settings & Alarms
. Set the Oxygen therapy controls as follows:
Control
Setting
Oxygen Source
External High-Pressure
Oxygen Delivery Mode
FiO2
FiO2
<desired setting>
Low and High FiO2 Alarms
<desired settings>
6
Press
Exit
to return to the
Oxygen
screen, and then begin using the Preset you configured for Oxygen therapy.
Press the Preset to highlight it, and then press
START
beneath it.
7
You are now ready to connect the patient and deliver High Flow therapy.