Velp Scientifica BOD Sensor F102B0133 Instruction Manual Download Page 3

 

1.  Introduction 

BOD  measurement  with  BOD  Sensor  is  based  on  the  respirometric  method  that  provides  a  direct  measurement  of  the 
oxygen  consumed  by  microorganisms  from  an  air  or  oxygen-enriched  environment  in  a  closed  atmosphere,  at  a  constant 
temperature (20 °C ± 0.5) and with a soft sample stirring. 

2.  Assembly and installation 

Upon  receipt  and  after  having  removed  the  packaging,  please  check  the  integrity  of  the  instrument.  The  box  includes:

  BOD Sensor 

  Alkali holder 

  Instruction manual 

 

BOD Sensor is delivered with 2 lithium batteries installed: remove the insulating strip from the rear without opening it. 

3.  Operating controls 

The display of the electronic measuring unit is normally turned off reducing he consumption of internal batteries. The touch 
of a key  activates the display which automatically  is turned off  after  three seconds. The key  activates different functions  if 
the unit is performing a measurement or if it is in memory code. 

3.1 Description of keys 

Out of a measuring cycle

 

(before starting) 

Key 

SET

 with display off 

 

 

The scale is shown 

Key 

SET

 with display on 

 

 

The scale is changed 

Key 

START

 with display off 

 

 

The scale is shown 

Key 

START

 with display on 

 

 

The cycle of measurement is started 

During a measuring cycle 

(after starting)

 

Key 

SET

 

 

 

 

 

The scale is shown 

Key 

START

 

 

 

 

 

The current measure is shown 

(also after the 5-day period if Reset is not performed) 

Key 

START

 kept pressed 

 

 

The memory mode is entered 

(also after the 5-day period if Reset is not performed) 

Memory mode 

(it is active during a measuring cycle and also after if Reset is not performed) 

Key 

SET

 

 

 

 

 

The stored value is shown 

Key 

START

 

 

 

 

 

The order number of stored value is increased 

If no values are stored, the display shows three points after a prolonged pressing of key START. 
Reset 
Keys 

SET

+

START

 kept pressed 

 

Measurement cycle stopped, stored values deleted and the scale is shown 

Before starting a new measuring cycle, it is requested to reset the stored values. 

3.2 Scale selection 

Scale 

Volume of sample 

The scale is related to the 

expected BOD value. 

The possibility of measuring 

diluted samples extends the range.

 

A: 0 ÷ 999 mg O

2

/l 

100 ml 

B: 0 ÷ 600 mg O

2

/l 

150 ml 

C: 0 ÷ 250 mg O

2

/l 

250 ml 

D: 0 ÷ 90 mg O

2

/l 

400 ml 

3.3 Alarms 
---

 

The measured value overcomes the maximum value of the chosen scale.  

bAt

 

Battery replacement. The charge of batteries does not assure in this case the finishing of a complete cycle.  

4.  Maintenance 

BOD Sensor warns about the required change of batteries when the unit is resetted and the charge is lower than a minimum 
threshold. Using a coin lever in the rear part of BOD Sensor and open it. Batteries are removed using the fingers. Introduce 
two fresh batteries at the same time paying attention to the polarity. Battery replacement deletes any stored value. 

5.  Technical data 

Dimensions (WxHxD) 

50x70x70 mm 

Weight 

80 g 

Measurement 

Electronic pressure probe 

BOD values 

Directly on display at any time 

Stored data 

5 BOD values at 24 hours 

Display 

3 LED digits, 7 mm high 

Precision 

± 1 digit (

 ±3.55 hPa) ± 1% 

Operation range 

500 ÷ 1100 mbar (h Pa) 

Scales 

90, 250, 600, 999 ppm BOD 

Power supply 

2 lithium batteries (250 mAh) 

Electrical input with display on  15 mA max. 

Electrical input with display off  7,5 µA 

Protection degree 

IP 54-EN 60529 

Admitted room temperature 

Storing 

–25 

 +65° C               

Ope5 

 +40°C 

6.  Accessories / Spare parts 

A00000135 

BOD Sensor Check 

 

 

A00000136 

Control Test Tablets  

10000039 

Battery BOD Sensor 

 

 

10000904 

Alkali container 

EN 

Summary of Contents for BOD Sensor F102B0133

Page 1: ...lo debe utilizarse para aplicaciones de laboratorio El fabricante declina toda responsabilidad por el uso no conforme a las instrucciones de los dispositivos Dieses Ger t darf nur f r Laboranwendungen...

Page 2: ...of sample according to adopted scale measured by a graduated cylinder 2 Put into each bottle a magnetic stirring bar 3 Fill the alkali holder small container located under bottle cap with an amount o...

Page 3: ...ART The order number of stored value is increased If no values are stored the display shows three points after a prolonged pressing of key START Reset Keys SET START kept pressed Measurement cycle sto...

Page 4: ...ra effettuato nessuna memorizzazione alla pressione prolungata del tasto START sul display si illuminano tre punti stando ad indicare che BOD Sensor non ha ancora memorizzato nessun dato Reset Tasto S...

Page 5: ...n a rien enregistr dans la m moire appuyez e maintenez le bouton START sur l cran s allume trois points indiquant que la m moire est vide Remise z ro Touche SET START simultan ment Arr te le cycle de...

Page 6: ...mbiar el n mero de la medici n almacenada Si BOD Sensor no ha hecho ning n almacenamiento al pulsar del bot n START en la pantalla se encender tres puntos indicando que ning n dato es en la memoria Re...

Page 7: ...das Display drei Punkte nach einem l ngeren Dr cken der Taste START Reset Tasten SET START gleichzeitig Messung wird gestoppt Speicher wird gel scht und die Skala wird angezeigt Zuerst der Analyse beg...

Page 8: ...yzers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e M...

Reviews: