5.4 Resistance
Ranges :
x 1 - 0.2
Ω
up to 2 k
Ω
, midscale, at 20
Ω
x 10 - 2
Ω
up to 20 k
Ω
, midscale, at 200
Ω
x 100 - 20
Ω
up to 200 k
Ω
, midscale, at 2 k
Ω
x 1k - 200
Ω
up to 2 M
Ω
, midscale, at 20 k
Ω
x 10k - 2 k
Ω
up to 20 M
Ω
, midscale, at 200 k
Ω
Accuracy at FSD : 3
5.5 General
Iceo 150 µA - 1.5 mA - 15 mA - 150 mA
hFE 0 -1000 (with extra connector)
Size : 148 x 100 x 35 mm
Weight : approx. 280g
6. Accessories
Test leads
Battery (9V)
Operating manual
7. Battery and fuse replacement
Remove screw on the back cover and open the case. Replace the exhausted
battery with a new one.
Fuse rarely needs replacement and blows usually as a result of the operator's error.
Open the case as mentioned above and replace the blown fuse with ratings
specified : F 1.5 A / 250 V
Remark :
Before attempting to open the case, be sure that the test leads have been
disconnected from measurement circuit to avoid electric shock.