6
©
Velleman
HPS40 Technical Doc
DATE REVISION
OVERVIEW
07-01-2003
Value R92 changed from 33K to 22K for improved RS232 com.
27-01-2003
IL211AT replaced by IL213AT improved RS232 communication
27-01-2003
R26 changed from 270 to 470, improved oscillator stability for
oversampling