2.2. ELECTROMAGNETIC EMISSION
RF emission CISPR 11, Group 1, Class A
Harmonic Emissions IEC 61000-3-2, Class A
Voltage changes/fluctuations and flickers IEC 61000-3-3
NOTE:
The EMISSIONS specifications for this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 Class A).
If the chair is used in a residential environment (which normally requires CISPR 11 Class B), this equipment might not provide
adequate protection for radiofrequency communication services. The user might have to take necessary precautions,
such as moving or adjusting the equipment.
Immunity Test
Test Level Professional Healthcare
Electrostatic discharge (ESD) IEC 61000-4-2
± 8 kV contact disharge
± 2,4,6,8,15 kV air discharge
Radiated RF field IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80% AM 1 kHz
Proximity fields from wireless transmitters
IEC 61000-4-3
80 MHz to 2700Hz. 3V/m
Spot Test: 385 MHz. at 27V/m;
(710,745,780,5240,5500,5785) MHz. at 9V/m;
(450,810,870,930,1720,1845,1970,2450) MHz. At 28V/m
Electrical fast transient / burst IEC 61000-
4-4
± 2kV, AC mains
± 1kV, I/0 ports
100 kHz PRR
Surge IEC 61000-4-5
AC mains, Line to Ground
AC mains, Line to Line
± 0.5, 1, 2 kV
± 0.5, 1 kV
Conducted RF IEC 61000-4-6
3 V (0.15 – 80 MHz)
6 V ISM Bands
80% AM 1 kHz
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m – 50 or 60 Hz
Voltage dips, short interruptions and voltage
variations on power supply input lines IEC
61000-4-11
100% dip, 0.5 periods, 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
100% dip, 1 period 30% dip,
25/30 periods (50/60 Hz) Interrupt 100% drop, 5 sec.