background image

3

1 For your safety

VEGASON 63 • Four-wire 4 … 20 mA/HART

51864-EN-210616

1  For your safety

1.1  Authorised personnel

All operations described in this documentation must be carried out 

only by trained, qualified personnel authorised by the plant operator.

During work on and with the device, the required personal protective 

equipment must always be worn.

1.2  Appropriate use

VEGASON 63 is a sensor for continuous level measurement. 

You can find detailed information about the area of application in 

chapter " 

Product description

". 

Operational reliability is ensured only if the instrument is properly 

used according to the specifications in the operating instructions 

manual as well as possible supplementary instructions.

1.3  Warning about incorrect use

Inappropriate or incorrect use of this product can give rise to applica-

tion-specific hazards, e.g. vessel overfill through incorrect mounting 

or adjustment. Damage to property and persons or environmental 

contamination can result. Also, the protective characteristics of the 

instrument can be impaired.

1.4  General safety instructions

This is a state-of-the-art instrument complying with all prevailing 

regulations and directives. The instrument must only be operated in a 

technically flawless and reliable condition. The operator is responsi

-

ble for the trouble-free operation of the instrument. When measuring 

aggressive or corrosive media that can cause a dangerous situation 

if the instrument malfunctions, the operator has to implement suitable 

measures to make sure the instrument is functioning properly.
The safety instructions in this operating instructions manual, the na-

tional installation standards as well as the valid safety regulations and 

accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device 

beyond that described in the operating instructions manual may be 

carried out only by personnel authorised by the manufacturer. Arbi-

trary conversions or modifications are explicitly forbidden. For safety 

reasons, only the accessory specified by the manufacturer must be 

used.
To avoid any danger, the safety approval markings and safety tips on 

the device must also be observed.

1.5  EU conformity

The device fulfils the legal requirements of the applicable EU direc

-

tives. By affixing the CE marking, we confirm the conformity of the 

instrument with these directives.
The EU conformity declaration can be found on our homepage.

Summary of Contents for VEGASON 63

Page 1: ...Quick setup guide Ultrasonic sensor for continuous level measurement VEGASON 63 Four wire 4 20 mA HART Document ID 51864...

Page 2: ...ule PLICSCOM 9 5 1 Insert display and adjustment module 9 5 2 Setup steps 9 5 3 Menu schematic 12 6 Set up with smartphone tablet PC notebook via Bluetooth 14 6 1 Preparations 14 6 2 Connecting 15 6 3...

Page 3: ...plying with all prevailing regulations and directives The instrument must only be operated in a technically flawless and reliable condition The operator is responsi ble for the trouble free operation...

Page 4: ...ations are accepted as the standard in field instrumentation The device fulfils the requirements of the following NAMUR recom mendations NE 21 Electromagnetic compatibility of equipment NE 43 Signal l...

Page 5: ...pprovals process seal process temperature signal output voltage supply protection protection class Data matrix code for VEGA Tools app SIL identification with SIL rating ex works Move to www vega com...

Page 6: ...the so called dead zone in which measurement is not possible is maintained The exact value of the dead zone is stated in chapter Technical data 1 2 Fig 1 Minimum distance to the max level 1 blocking...

Page 7: ...e wire ends into the open terminals according to the wir ing plan Fig 3 Connection steps 5 and 6 7 Press down the opening levers of the terminals you will hear the terminal spring closing 8 Check the...

Page 8: ...pply VEGASON 63 Four wire 4 20 mA HART 51864 EN 210616 4 2 Wiring plan double chamber housing 4 20 mA PE L N L1 N GND 1 2 1 2 3 4 4 20mA IS Fig 4 Wiring plan double chamber housing 1 Voltage supply 2...

Page 9: ...positions in 90 offset can be selected and turn to the right until it clicks into place 3 Screw housing lid with inspection window tightly back on Disassembly is carried out in reverse order The disp...

Page 10: ...plica tion for example Aqueous solution Medium Liquid 4 Select in the menu item Application the vessel the application and the vessel form for example Storage tank Vessel shape Storage tank The sensor...

Page 11: ...and weld joints on the vessel walls cause interfering reflections which can impair the measurement A false echo storage detects and marks these false echoes so that they are no longer taken into accou...

Page 12: ...Depending on the version and application the highlighted menu windows may not always be available Basic adjustment Basic adjustment 1 Display Diagnostics Service Info Min adjustment 1 1 0 00 4 000 m...

Page 13: ...djustment 4 Display Diagnostics Service Info False signal suppression 4 1 Change now Extended setting 4 2 None Current output 4 3 Characteristic 4 20 mA Fail mode 3 6 mA Min current 3 8 mA Simulation...

Page 14: ...tooth 1 Switch On Bluetooth active Off Bluetooth not active The security concept of Bluetooth operation absolutely requires that the default setting of the sensor PIN be changed This prevents unau tho...

Page 15: ...PIN is still effective Note Bluetooth access can only be established if the current sensor PIN differs from the default setting 0000 It is possible both when the PIN is deactivated and when it is acti...

Page 16: ...essage Con necting is displayed For the first connection the operating device and the sensor must authenticate each other After successful authentication the next con nection functions without authent...

Page 17: ...ween lower edge of the transducer and medium surface Measuring range Liquids up to 15 m 49 21 ft Bulk solids up to 7 m 22 97 ft Dead zone 0 6 m 1 969 ft Electromechanical data version IP66 IP67 Cable...

Page 18: ...18 Notes VEGASON 63 Four wire 4 20 mA HART 51864 EN 210616...

Page 19: ...19 Notes VEGASON 63 Four wire 4 20 mA HART 51864 EN 210616...

Page 20: ...ing scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change withou...

Reviews: