![Vega VEGAPULS 66 enamel Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/vega/vegapuls-66-enamel/vegapuls-66-enamel_operating-instructions-manual_3984686033.webp)
33
6 Set up with the display and adjustment module PLICSCOM
VEGAPULS 66 enamel • 4 … 20 mA/HART - four-wire
31954-EN-141215
Basic adjustment
▶
Basic adjustment
1
Display
Diagnostics
Service
Info
Min. adjustment
1.1
0.00 %
=
30.000 m(d)
0.665 m(d)
Max. adjustment
1.2
100.00 %
=
0.000 m(d)
0.665 m(d)
Medium
1.3
Liquid
▼
not known
▼
Vessel form
1.4
not known
▼
Damping
1.5
0 s
Linearisation curve
1.6
linear
▼
Sensor-TAG
1.7
Sensor
Display
Basic adjustment
2
▶
Display
Diagnostics
Service
Info
Displayed value
2.1
Scaled
Display unit
2.2
Volume
▼
l
▼
Scaling
2.3
0 % = 0.0 l
100 % = 100.0 l
Backlight
2.4
Switched off
▼
Diagnostics
Basic adjustment
3
Display
▶
Diagnostics
Service
Info
Peak value indicator
3.1
Distance min.: 0.234 m(d)
Distance max.: 5.385 m(d)
Meas. reliability
3.2
15 dB
Sensor status
OK
Curve selection
3.3
Echo curve
▼
Echo curve
3.4
Presentation of the echo
curve